| Comet T0213 Specifications |
| Technical Parameters |
Value |
| Output |
0-10V |
| Measured Value |
Temperature + Relative Humidity |
| Construction Type |
Duct Mount (150mm) |
| Design |
Industrial |
| Relative humidity range |
0 to 100% |
| Accuracy of relative humidity measurement |
±2.5% relative humidity from 5 to 95% at 23°C |
| Accuracy of temperature output |
±0.4°C from -30 to +100°C
±0.4% from reading over +100°C |
| Available temperature units |
degrees Celsius, Fahrenheit |
| Accuracy and range of dew point temperature output - for more details see graphs |
±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80 °C |
| Accuracy and range of absolute humidity output |
±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3 |
| Accuracy and range of specific humidity output |
±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg |
| Accuracy and range of mixing ratio output |
±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg |
| Accuracy and range of specific enthalpy output |
±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg |
| Temperature operating range of case with electronics |
-30 to +80°C |
| Temperature operating range of LCD display |
readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
Range of humidity sensor temperature compensation all temperature range |
| Voltage outputs |
0-10V, outputs are not galvanic isolated - have common ground |
| Configuration of outputs and output range |
user adjustable from the PC |
| Filtering ability of sensor cover |
0.025mm - filter with stainless steel mesh |
| Protection of the case with electronics |
IP65 electronics, IP40 sensors |
| Power |
15-30Vdc, maximum consumption 20mA; 24 Vac |
| Dimensions |
display unit 88.5 x 250 x 39.5 mm (W x H x D), stem length 150 mm |
| Weight |
approximately 225g |
| Warranty |
3 years |