Test Voltages |
500 V – 1 kV - 5 kV – 10 kV directly, one button selectable
500 V to 10 kV in 25 V, 100 V or 500 V steps DC, negative |
Max Resistance Reading |
10 TΩ @ 10 kV |
DC Voltmeter |
15 V up to 600 VDC
Accuracy: ±(5% of reading + 3 digits) |
AC Voltmeter |
15 V up to 600 V RMS
Accuracy: ±(5% of reading + 3 digits) |
Overvoltage Category |
CAT. III - 600 V |
Leakage Current Measurement |
1 nA up to 1500 µA
±(10% of reading + 3 digits) |
Capacitance Measurement |
50 nF up to 10 µF @ 500 V
50 nF up to 5 µF @ 1000 V
30 nF up to 2 µF @ 2500 V
30 nF up to 1 µF @ 5000 V
30 nF up to 680 nF @ 10,000 V
Accuracy: ±10% of reading ± 3 digits |
Short Circuit Current |
Maximum 2 mA |
Test Voltages Accuracy |
±3% of nominal test voltages on 10 GΩ |