Earth Continuity
Test Current |
>200mA |
Display Range |
0.00ohms - 199ohms |
Resolution |
0.01ohms max |
Test Lead |
Compensation: 0.00ohms - 10.00ohms |
Live Circuit Detection |
Test inhibit if >30V detected |
Insulation Resistance
Test Voltage |
100V, 250V, 500V |
Display Range |
0.05Mohms - 199Mohms |
Resolution |
0.01Mohms maximum |
Live Circuit Detection Test |
inhibit if >30V detected |
Loop/Line Impedance
Supply Voltage |
195 - 253V/440V, 45Hz - 65Hz |
Test Current |
15mA (non tripping for RCD protected circuits) |
Display Range |
0.01ohms - 2000ohms |
Resolution |
0.01ohms maximum |
Loop Impedance |
Phase to Earth |
Line Impedance |
Phase to Neutral/Phase to Phase |
PFC/PSC Range |
0 - 26kA |
RCD Trip Time
Supply Voltage |
195 - 253V, 45 - 65Hz |
Test Currents |
10, 30, 100, 300, 500mA |
Trip Time Ranges |
0ms - 2000ms at ½IDn
0ms - 300ms at IDn general |
|