Earth continuity / resistance measurement
Display range |
0.00Ω to 199Ω |
Measurement range |
0.01Ω to 199Ω |
Accuracy |
±(2% rdg + 5d) |
Resolution |
0.01Ω maximum |
Open circuit test voltage |
4VDC, nominal |
Test leads zero |
Zero up to 10 Ω, by Zero button |
Number of measurements |
5,000 1 second tests |
Audible / visible warning |
≥ 30VAC/DC at inputs |
Insulation resistance (auto short circuit test)
Display range |
0.05MΩ to 200MΩ |
Measurement range |
0.05MΩ to 200MΩ |
Accuracy |
±(5% rdg + 5d)
0.05MΩ to 200MΩ
±(10% rdg +5d)
101MΩ to 200MΩ |
Resolution |
0.1MΩ maximum |
Open circuit test voltage |
250, 500, 1000V (as per IEC 61557-2) |
Short circuit test current |
<2mA |
Number of measurements |
5,000 1 second tests |
Audible / visible warning |
≥ 30VAC/DC at inputs |
User protection |
Test inhibited if ≥ 30VAC/DC at inputs |
Insulation resistance (point to point)
Display range |
0.05MΩ to 300MΩ |
Measurement range |
0.05MΩ to 300MΩ |
Accuracy |
±(5% rdg + 5d) |
Resolution |
0.01MΩ maximum |
Open circuit test voltage |
250, 500, 1000V (as per IEC 61557-2) |
Short circuit test current |
<1mA |
Number of measurements |
5,000 x 1 second tests |
Audible / visible warning |
≥ 30VAC/DC at inputs |
Circuitry protection |
Test inhibited if ≥ 30VAC/DC at inputs |
|