            Voltage measurement range | 
        
        
            |   | 
              | 
              | 
            1 mV | 
            10 mV | 
            100 mV | 
            1 V | 
            10 V | 
        
        
              | 
            1 µA | 
            Full scale resistance | 
            1 kΩ | 
            10 kΩ | 
            100 kΩ | 
            1 MΩ | 
            10 MΩ | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.2% | 
            0.06% | 
            0.06% | 
            ** | 
            ** | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            ** | 
            ** | 
        
        
            | 10 µA | 
            Full scale resistance | 
            100 Ω | 
            1 kΩ | 
            10 kΩ | 
            100 kΩ | 
            1 MΩ | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.2% | 
            0.06% | 
            0.06% | 
            0.06% | 
            ** | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.001% | 
            ** | 
        
        
            | 100 µA | 
            Full scale resistance | 
            10 Ω | 
            100 Ω | 
            1 kΩ | 
            10 kΩ | 
            100 kΩ | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.2% | 
            0.2% | 
            0.06% | 
            0.06% | 
            0.06% | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.5% | 
        
        
            | 1 mA | 
            Full scale resistance | 
            1 Ω | 
            10 Ω | 
            100 Ω | 
            1 kΩ | 
            10 kΩ | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.2% | 
            0.2% | 
            0.2% | 
            0.06% | 
            0.06% | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.5% | 
        
        
            | 10 mA | 
            Full scale resistance | 
            0.1 Ω | 
            1 Ω | 
            10 Ω | 
            100 Ω | 
            1 kΩ | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.5% | 
            0.2% | 
            0.2% | 
            0.2% | 
            0.2% | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.5% | 
        
        
            | 100 mA | 
            Full scale resistance | 
            0.01 Ω | 
            0.1 Ω | 
            1 Ω | 
            10 Ω | 
            100 Ω | 
        
        
            | Accuracy (1 year) calibration temperature °C ± 5 °C ± % reading1 | 
            0.5% | 
            0.2% | 
            0.2% | 
            0.2% | 
            ** | 
        
        
            | Temperature coefficient/°C 10 °C to 35 °C ≤ 65% RH non-condensing ± % reading (typical) | 
            0.001% | 
            0.001% | 
            0.001% | 
            0.001% | 
            ** |