Thiết bị mô phỏng kiểm tra điện từ trường EMC 3CTEST RWPN 6916TUL
Technical Parameters
|
Test Voltage
|
Max 6 kV
(the actual output depends on the set value of the generator)
|
Phase Sync.
|
Any L-L, L-PE at any phase angles between L1, L2, L3, N, PE
|
Coupling Paths
|
Any two lines between L1, L2, L3, N, PE
|
Coupling Path Switching Method
|
Automatic
|
Coupling Capacitor
|
3 μF
|
Coupling Resistor
|
5Ω, 12Ω, 27.5Ω, 30Ω
(use standard types or user-defined, depending on the internal impedance of the generator)
|
Coupling Attenuation
|
<2 dB
|
Residual Pulse Voltage At The EUT Injection Terminal
|
<15% of test voltage or rated peak voltage of the CDN
|
General Parameters
|
Operating Power Supply
|
AC 110 V / 220 V ±10%, 50/60 Hz ±5%
(Default setting as AC 220 V, 50 Hz for Mainland China)
|
Fuse
|
6 A
|
Max Power
|
200 W
|
Auxiliary Port
|
D-sub 25p
|
Operating Status Indicators
|
Front panel LED
|
Grounding Method
|
Grounding strap
|
Dimension
|
19”, 6 U
|
Weight
|
Approx 30 kg
|
Operating Temp.
|
15℃ - 35℃
|
Operating Humidity
|
45% - 75%
|
Operating Air Pressure
|
86 kPa – 106 kPa
|
|