Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Continuous Wave Simulator, 80W CWS 500N1.3 EM TEST

Mã sản phẩm: CWS 500N1.3
Sử dụng cho Model:
Hãng SX: EM TEST
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng
The CWS 500N1.3 is the most compact single box test equipment for testing conducted rf immunity per IEC 61000-4-6 Ed. 3 and Ed. 4 and related standards with a frequency range of 10 kHz to 400 MHz. Apart from the 1 kHz 80 % AM signal the generator also generates a 2 Hz 80 % AM signal to test medical appliances and a 1 Hz PM signal with 50 % duty cycle required to test safety equipment like fire alarms.

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Highlights

Test procedure acc. IEC 61000-4-6 Ed. 4
Built-in amplifier saturation and linearity check
Built-in amplifier, 10 kHz to 400 MHz
Built-in power meter, 9 kHz to 1 GHz
Built-in dual directional coupler
Automatic calibration via built-in power meter
Connection for external amplifiers up to 1 GHz
USB-Interface
Software for Windows 7, Windows 8


Benefits:

One single box replaces five dedicated items. Hence setting up for testing is much easier and faster, saves space, is much more reliable and you achieve a higher reproducibility of the test results. 


The CWS 500N1.3 offers the best benefit-to-cost ratio. Calibration/levelling of the test set-up has never been easier; terminate the DUT port with the calibration adapter, connect this to the 150ohm-to-50ohm matching impedance and lead its output to the Cal Input. Start the calibration routine and you are done in a minute.


The signal generator delivers signals up to 400MHz for the internal amplifier while the frequency range can be extended for 9kHz to 1GHz using an external amplifier.
The powerful, top-performance built-in class A amplifier allows to test up to 30V using any kind of coupling/decoupling device (CDNs, EM clamps, current injection clamps). The built-in 1GHz current monitor offers the ability to control the injected current signal while using injection clamps.

Bình luận