HF-Analyser
Model : HFW59D
Hãng : Gigahertz

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

- FULL 10 GHz.
- Maximum video band width captures even the shortest radar pulses (< 0.5 µs).
- Optimized Peak Hold function (patent no. DE 103 34 886).
- 1 µW/m² display resolution (for higher sensitivity, e.g. for wireless LAN signals, use HV20_2400G10).
- Includes an especially shielded log-per antenna with excellent directional characteristics.
- Integrated steep high pass filter at 2.4 GHz reliably suppresses the often dominant GSM and DECT frequencies.
- Calibrated AC measurement output for the demodulated signal, scalable DC output.
- The meter is especially designed for peak measurements of modulated signals ("pulse" mode).

Thông số kỹ thuật

- Frequency range: 2.4 - 10.0 GHz
- Measurement range: Power flux density: 1 - 1999 µW/m²
- Precision:
   + Basic accuracy including linearity tolerance: +/- 3 dB 
   + Zero offset and rollover +/- 5 digits
- Sensor: Optimised logarithmic periodic antenna with frequeny compensation directly on the antenna, better directionality (patent pending AZ 10 2006 051 180.8)
- Audio analysis: Identification of pulsed radiation sources (WLAN, WiFi, radar, etc.) and determination of increased contamination supported by an acoustic signal proportional to the modulation frequency and the field strength
- Signal rating: Display of peak value as well as average value of the modulated signal part (in practice corresponds to the total signal). Please see manual for the evaluation of crest factors. - Measurement of CW-(test-)signals not possible.
- Signal output ports: AC (demodulated signal, calibrated) and scalable DC output.
- Power supply:
   + Rechargeable highpower 9.6 Volt NiMH battery pack inside the meter, average operation time: 7 - 8 hours
   + Low-Batt. indication
   + Auto-power-off function (automatically deactivated for longterm measurements)
   + AC-adaptor for charging and longterm operation included