- Solar modules test
- Impulse voltage: 10 kV
- Waveform: 1,2/50µs
- Energy: 250 J
- Cp = 0 - 183 nF
- The HV-Pulse generator PG10-200 / PG12-360 is used for impulse voltage tests of solar modules (photovoltaic panels) with the standard surge voltage waveform 1.2/50 μs according to IEC 60060-1/2 up to 10 kV acc. to. IEC 61730-1/2 / EN 61730-1.
- For the surge voltage tests of solar modules, these are enveloped with a copper foil as described in the relative above standard. After that the connections of the solar modules are tested with surge voltage against the copper foils. The tight copper foil envelope required by the standard results in comparatively high capacitances of approx. 0 - 183 nF, which is switched in parallel to the output of the impulse generator.