Bronkhorst® model D-6310 Mass Flow Meters (MFMs) are suited for precise measurement of flow ranges between 10…200 mln/min and 0,1…2 ln/min at operating pressures between vacuum and 10 bar (Aluminium) or 20 bar (Stainless Steel). The MFM consists of a proven inline thermal (CTA) mass flow sensor and a microprocessor based pc-board with signal and fieldbus conversion and a PID controller for optional mass flow control by means of a separately mounted control valve. The instument is IP65 complient and can optionally be equipped with a modern, multi-functional and multi-colour display, with operator buttons on the instrument.
The digital MASS-STREAM™ series is characterized by a high degree of signal integrity and, as an option, up to 8 calibration curves of different gases and process conditions can be memorized in the instrument. In addition to the standard RS232 output the instruments also offer analog I/O. As an option, an on-board interface can be mounted to provide DeviceNet™, PROFIBUS DP, Modbus-RTU or FLOW-BUS protocols.
Proven direct inline CTA sensor (no bypass)
Compact, rugged design (IP65, dust and waterproof)
Very low pressure drop
Less sensitivity to humidity or dirt
Optional with integrated TFT display
Thông số kỹ thuật
Measurement / control system |
Accuracy (based on calibration with Air at 5 bar (a) and T = 20 °C) |
±2% FS including non-linearity*
±1,5% FS on request* |
Repeatability |
< 0,2% FS |
Turndown |
1:20 |
Type of gases |
almost all gases, compatible with chosen materials |
Response time sensor |
approx. 0,3 sec. |
Operating temperature |
0…50°C |
Temperature sensitivity |
±0.2% Rd/°C (Air) |
Pressure sensitivity |
±0,3% Rd/bar typical (Air) |
Leak integrity |
tested < 2 x 10-8 mbar l/s He |
Attitude sensitivity |
at 90° deviation from horizontal max. error 0,2 % at 1 bar typical N2 |
Warm-up time |
30 min. for optimum accuracy,
within 30 seconds for accuracy ±4% FS |