Features:
- Smaller and over 50% lighter than competitive instruments
- Water-resistant design in an integral concussion-proof enclosure
- Continuous LCD shows simultaneous gas concentrations for five gases
- 0 - 1000 ppm measurement of VOCs (Volatile Organic Compounds)
- Integral motorized pump option for remote sampling
- Two power options: AA alkaline or rechargeable hot-swappable battery pack
- Provides 95 dB alarm tone and two bright wide-angled alarm bars
- Equipped with internal vibrating alarm for high noise areas
- Multi-language support in English, French, German, Spanish, and Portuguese
- Three backlight options: standard- activates in low light (auto), in alarm (auto) and reactivates on-demand; alarm only- activates in alarm conditions and reactivates on-demand; stealth mode
- Four alarm levels: instantaneous Low and High Alarm for all gases; TWA (time-weighted average) and STEL (short-term exposure limit) for VOCs and toxic gases
- Simple automatic calibration procedure; compatible with BW MicroDock II automatic test and calibration station
- Records TWA, STEL, and maximum exposures to gas and displays readings on command
- User selectable field options include: PID correction factor, language selection, confidence beep, span level, set TWA/STEL calculation choice, set calibration due date, pass code protection, latching alarm, 'SAFE' display function, stealth mode, and selectable ppm resolution for some gases (e.g. SO2 0.1 ppm)
- Full function self-test of: sensor, battery and circuit integrity; and audible/visual alarms
Supported Gases:
- H2S - hydrogen sulfide
- CO - carbon monoxide
- O2 - oxygen
- HCN - hydrogen cyanide
- Cl2 - chlorine
- ClO2 - chlorine dioxide
- SO2 - sulfur dioxide
- O3 - ozone
- PH3 - phosphine
- NH3 - ammonia
- NO2 - nitrogen dioxide
- LEL combustibles
- PID - photo ionization
Description
- GasAlertMicro5 PID portable gas detector allows simultaneous monitoring of up to five atmospheric hazards including volatile organic compounds (VOCs), combustible gas (% LEL), oxygen, and a wide selection of individual electrochemical toxic sensors.
- Compact, lightweight, and rugged, the GasAlertMicro5 PID activates audible, visual and vibrating alarms in the event of a low, high, TWA or STEL alarm condition. VOCs can present multiple potential threats in the workplace environment.
- VOCs are characterized by their tendency to evaporate easily at room temperature and include substances such as paint thinners, benzene, solvents, butadiene, toluene, hexane, as well as vapors associated with fuels such as diesel, heating oil, gasoline, kerosene, and jet fuel.
- The GasAlertMicro5 PID’s integrated photoionization detector (PID) enables sensitive broad spectrum monitoring of volatile organic contaminants. Highly configurable to suit a range of applications, GasAlertMicro5 PID comes as either a diffusion or pumped instrument; with either NiMH rechargeable hot-swappable battery pack or normal alkaline and has optional datalogging capabilities.
Specifications
Gas
|
Measuring range
|
Resolution
|
TWA
|
STEL
|
LOW
|
HIGH
|
O2
|
0-30.0%
|
0.1
|
N/A
|
N/A
|
19.5
|
22.5
|
CO
|
0-999 ppm
|
1
|
35
|
200
|
35
|
200
|
H2S
|
0-100 ppm
|
1
|
10
|
15
|
10
|
15
|
PH3
|
0-5.0 ppm
|
0.1
|
0.3
|
1.0
|
0.3
|
1.0
|
SO2
|
0-100 ppm
|
0.1
|
2
|
5
|
2
|
5
|
NO2
|
0-99.9 ppm
|
0.1
|
2.0
|
5.0
|
2.0
|
5.0
|
HCN
|
0-30.0 ppm
|
0.1
|
4.7
|
10.0
|
4.7
|
10.0
|
Cl2
|
0-50.0 ppm
|
0.1
|
0.5
|
1.0
|
0.5
|
1.0
|
NH3
|
0-100 ppm
|
1
|
25
|
35
|
25
|
50
|
ClO2
|
0-1 ppm
|
0.1
|
0.1
|
0.3
|
0.1
|
0.3
|
O3
|
0-1 ppm
|
0.1
|
0.1
|
0.1
|
0.1
|
0.1
|
PID
|
0-999 ppm
|
5.0
|
50
|
100
|
50
|
100
|
LEL
|
0-100% LEL
0-5% v/v
|
1%
|
N/A
|
N/A
|
10%
|
20%
|
Auto zero
|
Yes
|
Calibration
|
Automatic
|
Backlight
|
Standard Mode: low light (auto); alarm condition (auto); reactivation on demand
|
Alarm Only Mode
|
alarm condition (auto); reactivation on demand
|
Stealth Mode
|
backlight inactive; audible and visual alarms silenced
|
Sensor
|
PID (VOC) 10.6 eV lamp; plug-in, electrochemical cell (toxic and oxygen); catalytic (LEL)
|
User field options
|
Confidence beep, set STEL period, force calibration on start-up, enable / disable sensor, pass code protection, latching alarms, span level, passed calibration user-lockout option, “SAFE†display function, stealth mode, fast pump, oxygen auto-cal on start-up, alarm only backlight mode, combustible gas measurement (% LEL or % by volume methane), LCD language option, STEL calculation choice, TWA calculation choice, PID / combustible correction factor options, user-settable calibration gas level, high resolution, clock
|
Pump
|
Tubing length up to 100 ft. / 30 m (50 ft. / 15 m standard) with user options
|
Alarm Indicators
|
Clearly advises alarm conditions with audible, visual, and vibrator alarms
|
Visual alarms
|
Two flashing alarm bars visible from all angles
|
LCD
|
indicates gas present and alarm levels encountered
|
Audible alarm
|
Typically 95 dB at 1 ft. / 30 cm variable pulsed audible alarm
|
Vibrator alarm
|
Pulses warning in gas alarm conditions, or for any status alarms
|
Status alarms
|
Low battery, over range, multiple gas / pump and failed or missing sensor
|
ALARM LEVELS
|
Instantaneous Low Alarm for all gases; Instantaneous High Alarm for all gases
TWA (time-weighted average) and STEL (short-term exposure limit) for toxics/VOCs
|
LCD
|
Continuous, alphanumeric gas readout and status display advises:
|
Gases monitored
|
VOCs (PID), H2S, CO, O2, SO2, PH3, NO2, NH3, HCN, Cl2, ClO2, O3 and/or combustibles (0-100% LEL or 0-5.0% methane)
|
Gas concentration
|
Simultaneously and continuously displays gas concentration(s) for all sensors (in ppm, % LEL, or % v/v)
|
Peak (max)/STEL/ TWA ppm or %
|
Records exposures and displays on demand
|
Alarm setpoints
|
Displayed on activation and on demand
|
Calibration due date
|
Days remaining on start-up
|
Clock
|
Real-time clock setting displayed on start-up
|
Tests
|
Full function self-test
|
Sensor integrity, circuitry, battery and audible/visual alarms on activation
|
Battery
|
Continuous
|
General Specifications
|
Rugged, composite material; two-shot molded polymer case with integral anti-shock boot
|
Operating temperature
|
14 to 104oF / -10 to 40oC VOC (PID); -4 to 122oF / -20 to 50oC other gases
|
Humidity
|
0 - 95% RH (non-condensing)
|
EMI/RFI
|
Complies with EMC Directive 89/336/EEC
|
Protection
|
Highly water resistant and dust proof housing
|
Battery Options
|
12 hours on three AA alkaline cells or 10 hours on rechargeable hot-swappable battery pack when equipped with PID, oxygen, combustible and TwinTox (H2S/CO) sensor
|
Size
|
5.7 x 2.9 x 1.5 in. / 14.5 x 7.4 x 3.8 cm
|
Weight
|
13.1 oz. / 370 g
|
Warranty
|
Full 2-year warranty including all sensors (1 year PID lamp and NH3 sensor)
|
Datalogger Options
|
User-downloadable datalogger
|
Data recorder
|
All events and occurrences (calibration zero adjustment, calibration span adjustment, gas readings, TWA, STEL, alarm condition, failures detected by the microprocessor).
|
Sampling rate
|
Factory set at 5-second intervals and user-settable from 1 to 127-second intervals
|
Storage
|
2 months of continuous data at 5 second intervals via 32 MB Flash MMC card (64 MB and 128 MB cards also available); when full, the wrap-around feature replaces oldest data with most recent data
|
|