Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Máy đo từ thông, từ trường Lake Shore Model 480 Fluxmeter

Mã sản phẩm:
Sử dụng cho Model:
Hãng SX:
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

Máy đo từ thông, từ trường Lake Shore Model 480 Fluxmeter

Liên hệ hỗ trợ: 0984.843.683.

Email: hien@ttech.vn

Thông tin sản phẩm

Máy đo từ thông, từ trường Lake Shore Model 480 Fluxmeter


Measurement

Number of units: 1

Input type: 2-lead, ground referenced

Input resistance: 100 kΩ or 10 kΩ

Maximum operating input voltage: 60 V

Absolute maximum input voltage: 100 V — WARNING — voltages between 60 V and 100 V will not damage the instrument but could result in personal injury or damage to other instruments

Update rate: 5 rdg/s display; 30 rdg/s IEEE-488; 30 rdg/s serial

DC

DC resolution: To 5¾ digits

DC integrator capacitance: 1 µF nominal

DC input resistance: 100 kΩ 10 kΩ

DC ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs

DC resolution: 0.001 mVs, 0.0005 mVs, 0.0005 mVs, 0.0005 mVs

DC accuracy: Offset: ±10 µVs ±DC integrator drift; Gain: ±0.25% of reading (<10 Vs/s maximum rate of change)

DC minimum dΦ/dt: 20 µVs/min

DC maximum dΦ/dt: 60 Vs/s

DC integrator drift: ±1 µVs/min, 0.0004% FS/min on 300 mVs range (100 kΩ input resistance constant temperature environment)

DC peak

DC peak resolution: 4¾ digits

DC peak integrator capacitance: 1 µF nominal

DC peak input resistance: 100 kΩ 10 kΩ

DC peak ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs

DC peak resolution: 0.01 mVs, 0.001 mVs, 0.001 mVs, 0.001 mVs

DC peak minimum reading: 0.05 mVs, 0.005 mVs, 0.005 mVs, 0.005 mVs

DC peak accuracy: Offset: ±100 µVs ±DC integrator drift; Gain: ±5% of reading (<10 Vs/s maximum rate of change)

DC peak maximum dF/dt: 60 Vs/s

DC peak update rate: Reduces update rate to ¼ normal

AC

AC resolution: 4¾ digits (reduced to 3¾ digits on the 30 µVs range)

AC integrator capacitance: 0.1 µF nominal

AC input resistance: 100 kΩ

AC ranges: 30 mVs, 3 mVs, 300 µVs, 30 µVs

AC resolution: 0.001 mVs, 0.0001 mVs, 0.01 µVs, 0.01 µVs

AC minimum rdg: 3.000 mVs, 0.3000 mVs, 30.00 µVs, 3.00 µVs

AC frequency response: 2 Hz to 50 kHz

AC accuracy: ±1% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±5% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)

AC integrator drift: NA

AC peak

AC peak resolution: 3¾ digits

AC peak integrator capacitance: 0.1 µF nominal

AC input resistance: 100 kΩ

AC peak ranges: 30 mVs, 3 mVs, 300 µVs

AC peak resolution: 0.01 mVs, 0.001 mVs, 1 µVs

AC minimum reading: 0.01 mVs, 0.001 mVs, 5 µVs

AC peak accuracy: ±5% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±10% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)

AC peak update rate: Reduces update rate to ¼ normal 

Front panel

Display type: 2-line × 20-character, vacuum fluorescent display

Display resolution: To ±5¾ digits

Display update rate: 5 rdg/s

Display units: Vs, MxN, WbN, VF, MxF, WbF, G, T, Wbcm, A, %

Units multipliers: p, n, µ, m, k, M, G

Annunciators: AC – AC input signal, DC – DC input signal, ∧ ∨ – Positive and negative peaks, R – Remote operation, :Alarm on

Keypad: 21 full travel keys

Interfaces

IEEE-488.2 capabilities: SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1

Serial interface: RS-232C electrical, DA-9 connector, 9600 BAUD

External reset type: Contact closure

Software support: LabVIEW™ driver

Alarms

Number: 2

Settings: High and low set point, Inside/Outside, Audible

Actuators: Display annunciator, beeper, relays

Relays

Number: 3

Contacts: Normally open (NO), normally closed (NC), and common (C)

Contact rating: 30 VDC at 2 A

Operation: Follow high, low alarms with third relay indicating no alarm state — can be operated manually

Connector: Detachable terminal block

Monitor analog output

Scale: ±3 V = ±FS on Vs range

Accuracy: ±1% of reading ±10 mV, (DC to 10 kHz); ±5% of reading ±10 mv, (10 kHz to 50 kHz)

Minimum load resistance: 1 kΩ

Connector: Detachable terminal block

Corrected analog output

Scale: User selected

Range: ±10 V

Resolution: 0.3 mV

Accuracy: ±2.5 mV

Minimum load resistance: 1 kΩ

Connector: Detachable terminal block/p>

General

Ambient temperature: 15 ºC to 35 ºC at rated accuracy; 5 ºC to 40 ºC with reduced accuracy

Power requirement: 100, 120, 220, 240 VAC, +5%, -10%, 50 or 60 Hz, 20 VA

Size: 216 mm W × 89 mm H × 318 mm D (8.5 in × 3.5 in × 12.5 in), half rack

Weight: 3 kg (6.6 lb)

Approval: CE mark, RoHS

Bình luận