Performance |
Operation range |
See Detection Capability table |
Detection limit (LDL, 3σ or 24 h peak-to-peak) |
See Detection Capability table |
Precision (1σ, greater of) |
± 0.75% or 1/3 of LDL |
Accuracy (greater of) |
± 4% or the LDL |
Zero drift |
None |
Speed of response |
2 minutes to 80% |
Environmental conditions |
10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature |
-10° to 50°C |
Gas Handling System and Conditions* |
Wetted materials |
Sample gas compatible |
Gas connections |
1/4" male VCR inlet & outlet |
Leak tested to |
1 x 10-9 mbar l /sec |
Inlet pressure |
0 – 15 psig (1 – 2 bara) |
Outlet pressure |
Vacuum (<10 Torr) |
Flow rate |
∼1 slpm |
Sample gases |
Ambient air, clean dry air (CDA), or N2 |
Gas temperature |
Up to 60°C |
Dimensions |
Standard sensor |
8.73" x 8.57" x 23.6" (222 mm x 218 mm x 599 mm) |
Sensor rack |
8.73" x 19.0" x 23.6" (222 mm x 483 mm x 599 mm) |
Weight |
Standard sensor |
33 lbs (15 kg) |
AMC cart (excl. sensors and pumps) |
260 lbs (118 kg) |
Electrical |
0-5 VDC, isolated 4-20mA or 0-20 mA output |
Alarm indicators |
2 user programmable, 1 system fault, Form C relays |
Power requirements |
90 – 240 VAC 50/60 Hz |
Power consumption |
40 Watts max. per sensor |
Signal output |
Isolated 4–20 mA per sensor |
User interfaces |
5.7" LCD touchscreen, 10/100 Base-T Ethernet, RS-232 |
|
802.11g Wireless (optional), Modbus TCP (optional) |
Patents |
0.2 ppb |