Performance |
Operating range |
See Detection Capability table |
Detection limit (LDL, 24 h peak-to-peak) |
See Detection Capability table |
Sensitivity (3σ) |
See Detection Capability table |
Precision (1σ, greater of) |
± 0.5% or 1/3 of Sensitivity |
Accuracy (greater of) |
± 4% or 1/2 of LDL |
Speed of response |
< 5 minutes to 95% (in 4-channel operation) |
Environmental conditions |
10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature |
-10°C to 50°C |
Gas Handling System and Conditions |
Wetted materials |
316L stainless steel, 10 Ra surface finish |
Gas connections |
1/4” male VCR inlet and outlet |
Leak tested to |
1 x 10-9 mbar l / sec |
Sample inlet pressure |
20 − 125 psig (2.4 − 9.6 bara) |
Flow rate |
< 1 slpm (gas dependent) |
Sample gases |
Inert gases, hydrogen and oxygen |
Gas temperature |
Up to 60°C |
Dimensions, H x W x D |
Standard sensor |
12.25" x 17.50" x 29.65" (311 mm x 445 mm x 753 mm) |
Weight |
Standard sensor |
70.5 lbs (32.0 kg) |
Electrical |
Alarm indicators |
User programmable setpoints (1 per channel), Form C relays |
Power requirements |
90 − 240 VAC, 50/60 Hz |
Power consumption |
400 Watts max. |
Signal output |
0−5 VDC analog, isolated 0−20 mA or 4−20 mA |
User interfaces |
10.4” LCD touchscreen, PS/2 for mouse and keyboard |
|
10/100 Base-T Ethernet, 4 USB ports, RS-232 |