Auto-Step Output Function |
n value sent automatically when n/m function selection is selected (two options: approximately 2.5 seconds/step or approximately 5 seconds/step) |
Sweep Functions |
Sweep time (two options: approximately 16 seconds or approximately 32 seconds) |
Memory Function |
50 value sets (generated and measured values are stored as value sets with the same address (up to 50 value sets can be stored)) |
Measuring Unit Maximum Input |
Voltage terminal: 300 V AC Current terminal: 120 mA DC |
Current Terminal Input Protection |
Fuses: 125 mA/250 V |
Measuring Unit Ground Voltage |
Maximum 300 V AC |
Measurement Display Updating Rate |
Approximately once per second |
Serial Interface |
Enabled when communication cable (RS232) is connected; sold separately as optional accessory |
Display |
Segmented LCD (Approx. 76 mm × 48 mm) |
Backlight |
LED backlight; auto-off after one minute (from when LIGHT key is turned on) |
Power Supply |
Four AA alkaline batteries, or special AC adapter (sold separately) |
Battery Life |
Measurement off, output 5 V DC/10 kO or greater: Approximately 40 hours Simultaneous signal generation/measurement, output 5 V DC/10 kO or greater: Approximately 20 hoursSimultaneous signal generation/measurement, output 20 mA/5 V: Approximately 12 hours (using alkaline batteries, with backlight off) |
Auto-Power-Off Function |
Approx. 10 minutes (auto-power-off can be disabled through a DIP switch setting) |
Applicable Standard |
EN55011: 1998, Class B, Group 1
EN61326-1: 1997 + A1: 1998
IEC61010-1, IEC61010-2-31 |
Insulation Resistance |
Across input terminal and output terminal, 500 V DC, 50 MO or greater |
Withstand Voltage |
Across input terminal and output terminal, 3.7 kVAC, for one minute |
Operating Temperature And Humidity Ranges |
0 - 50°C, 20 - 80% RH (no condensation) |
Storage Temperature And Humidity Ranges |
-20 - 50°C, 90% RH or less (no condensation) |
External Dimensions |
Approx. 190 × 120 × 55 mm |
Weight |
Approx. 730 g (including batteries) |