Power Requirements 24 to 28 VDC at 50 mA
Output Total load must equal 250 Ω ±5%
Cable O CI-*F-T5 in hazardous area: Two-conductor, twisted pair with connector; 22 AWG; 1,100 ft maximum cable length
O CG-*F-0 in nonhazardous (safe) area: Three-conductor; 22 AWG; 4,000 ft maximum cable length
Output: Max. load ( Ω) = [40 Ω × (PSV – 8)] – RC
Accuracy:
±1% of span at calibration point
±2% of span at the calibration point for the 0 to 10 ppm v range (316 stainless steel package only)