Application range |
up to 405 °C (°C/°F selectable) |
Ignition |
Gas and electric (interchangeable) |
Heating |
According to method or user-defined |
Stirring |
According to method or user-defined |
Cooling |
Built-in forced air (fan) |
Flash point detection |
Thermocouple |
Sample temperature |
Pt100 glass probe |
Sample temperature calibration |
Dynamic calibration against certified ASTM thermometer and
offset capability with correction table (21 calibration points) |
Barometric pressure correction |
Automatic correction |
Safety |
Overheat protection, automatic shut-off, fire-extinguishing system, password |
Interfaces |
3x USB, 1x RS232, 1x LAN |
Gas supply |
Test flame: 50 mbar of propane or butane
Fire extinguisher: 6 bar to 12 bar of nitrogen
or carbon dioxide |
Power supply |
115 V/230 V, 50 Hz/60 Hz, 1000 W |
Dimensions |
230 mm x 410 mm x 460 mm (W x D x H) |
Weight |
14 kg |