FEATURES |
|
■ |
Maximum sensitivity is ±5° full scale, and by offsetting it is possible to obtain direct readings to 0.1° regardless of the phase angle. |
■ |
DC voltage output proportional to the phase angle is available for operating a digital voltmeter when desired. |
■ |
Ease of operation and maintenance are the keynotes in design. |
|
Frequency Range |
10Hz to 2MHz |
Phase Angle Range |
0 to ±180°in six ranges |
±5°, ±10°, ±18°, ±50°, ±100°and ±180°full scale |
Input Voltage Range |
Two ranges: 5mV to 2Vrms and 100mV to 30Vrms |
(50mV to 20Vrms and 1 to 300Vrms at 10:1 probe used) |
Input Impedance |
1MΩ, 25pF in shunt |
Meter Offset |
in 10°steps to ±170° |
Meter Offset Accuracy |
±0.5% |
Meter Accuracy |
within ±[(2.5%+0.1°)+(offset error)] of full scale for each range |
Phase Output Signal |
10mV/degree, referred to the phase angle as measured |
Phase Output Signal Accuracy |
within ±(0.5%+0.1°) |
Phase Output Signal Response Time |
Approx. 30ms minimum |
Power Requirements |
AC 100V, 115V, 215V or 230V ±10% (Voltage selector provided)
50/60Hz; approx. 17VA
|
Dimensions Overall |
Approx. 430 (W)×165 (H)×360 (D) mm |
Weight |
Approx. 7kg |
Accessories, Furnished |
Input Cord : 2ea |
Output cord : 1ea |
|
|