Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Portable Analyzer for Carbon Dioxide for Agricultural Applications, Suitcase (K) Enclosure 302AK Nova Analytical Systems

Mã sản phẩm: 302AK
Sử dụng cho Model:
Hãng SX: Nova Analytical Systems
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng
Portable Analyzer for Carbon Dioxide for Agricultural Applications, Suitcase (K) Enclosure 
Model: 302AK
Hãng: Nova Analytical Systems

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

- The Model 302A Series Portable Gas Analyzers have been designed for accuracy, reliability, and ease of use and service. The 302A uses a solid state infrared detector which responds quickly to CO2 present in the gas sample.

- In operation, a built in sample pump draws in the gas sample through the sample hose, condensate removal filter, secondary filter and flow meter then on to the infrared CO2 detector. The detected CO2 is displayed on an LCD digital meter which has a switchable backlight for use in dark areas.

- A rechargeable battery provides enough power for about 8 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED tells when to recharge and a green LED verifies that it is receiving recharging power. The recharger is included.

- Rugged design that is easy to operate
- Fast warm-up and response
- Solid state infrared detector
- Digital readout meter with backlight
- Modular layout that is easy to maintain
- Rechargeable battery operation
- Built in sample pump, filter, and flowmeter
- Weatherproof (WP) cabinet with clear Lexan cover

Thông số kỹ thuật

- Method of Detection : Microprocessor based infrared detector for CO2
- Ranges Available : 0-3000, 0-5000 or 0-10,000 PPM; 0-2%, 0-5%, 0-10%, 0-20%, 0-50% or 0-100% CO2. Other ranges available.
- Resolution : 0.1% CO2 on % versions; 10 PPM CO2 on PPM versions
- Accuracy & Repeatability : Within ±2% full scale
- Drift : Within 1% full scale per 8 hours of continuous operation
- Response Time (T-90) : 30 - 40 seconds
- Ambient Temperature Range : 4 to 49 ºC; 40 to 120 ºF
- Linearity : Better than 1.0% of full scale
- Size & Weight : 14" L x 6" H x 10 1/2" D @ 12 lbs (35.5 x 15.2 x 26.6 cm @ 5.5 kg)
- Power : AC/DC Operation, 115 VAC 60 Hz for recharging
- Output Options : 4-20 mA or 0-1 VDC
- Alarms : High or low CO2 alarms with LED (optional)
- Hazardous Area Rating Options : General Purpose, Non-Hazardous