Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Portable Dew Point Analyzer 207K Nova Analytical Systems

Mã sản phẩm: 207K
Sử dụng cho Model:
Hãng SX: Nova Analytical Systems
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng
Portable Dew Point Analyzer
Model: 207K
Hãng: Nova Analytical Systems

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

- The Nova 207 Series Portable Dew Point Analyzers have been designed primarily for accurate dew point measurement of exothermic and endothermic heat treating atmospheres.
- The dielectric polymer sensor provides stable and accurate dew point measurements and is not affected by most gas compositions.
- A built in sample pump draws in samples that are of insufficient pressure from the process. The sample flow is shown on a small rotameter with valve. The analyzer then reads the dew point continuously.
- Standard range of -40 to +104°F (-40 to +40°C) Dew Point
- Rugged design that is easy to operate
- Use on any non-corrosive gas
- Long-life dielectric polymer sensor
- May be calibrated in the field
- Bright digital readout
- Built in sample pump, filter and flow meter
- Suitcase style (K) enclosure is standard
- Sensor easily replaced if contaminated
- Factory calibrated to NIST traceable standards

Thông số kỹ thuật

- Method of Detection : Dielectric polymer sensor that change capacitance with moisture variation
- Ranges Available : -40°F to +104°F (-40°C to +40°C) Dew Point
- Resolution :  1°F (1°C) Dew Point
- Accuracy & Repeatability : ±1°F (±1°C) Dew Point
- Drift : Less than 1% of full scale range per month
- Response Time (T-90) : Less than 30 seconds
- Ambient Temperature Range : -18 to 49 ºC; 0 to 120 ºF
- Linearity : Better than 1°F (1°C) Dew Point
- Size & Weight : K Suitcase Style - Approx. 14" L x 6" H x 11" D @ 12 lbs (36 x 15 x 27 cm @ 5.4 kg; WP style - approx. 12" L x 8" H x 7" D @ 8 lbs (30 x 20 x 18 cm @ 3.6 kg)
- Power : 115 VAC 60 Hz for recharging; 220VAC 50Hz available
- Output Options : 4-20 mA or 0-1 VDC
- Alarms : High or low dew point alarms with LED (optional)
- Hazardous Area Rating Options : General Purpose, Non-Hazardous

Bình luận