- The Nova 5000 Series Portable Flue Gas analyzer has been designed for accuracy, reliability, ease of use and ease of service, providing a detailed analysis of flue gas composition. The 5000 uses customer replaceable sensors which respond quickly to the O2, CO2, SO2, CO, and NOx (as NO) present in the flue gas sample. The O2 sensor life expectancy is typically 3-4 years. The CO, SO2, and NOx sensor life is typically 2-3 years. The infrared CO2 sensor should not need to be replaced under normal conditions of use.
- In operation, a built in sample pump draws in the flue gas sample through the stainless steel probe, 12ft. sample house, water separating filter, secondary filter, PTFE liquid blocker and flow meter then on to the gas sensors. The detected O2, CO2, SO2, CO, and NOx are displayed on separate large LCD meters with switchable backlights for use in dark areas. Optional with the Nova 5000 Series is a stack temperature reading.
- A rechargeable 'gel cell' battery provides enough power for about 4-5 hours of continuous operation and the analyzer can be used while it is being recharged. A red LED tells when to recharge and a green LED verifies that it is receiving recharging power. The Model 5000 series case is rugged, dust and water resistant when closed. A built in printer, which also shows date and time, is optionally available. A built in data logger is also available which can be uploaded to a PC.
- Rugged design that is easy to use
- Fast warm-up and accurate response
- Modular layout that is easy to maintain
- Digital readouts with switchable backlight
- Built in sample pump
- Rechargeable battery operation
- Active condensate removal
- Thermocouple temperature probe
- Long-life electrochemical sensors for O2, CO, SO2, and NOx
- Wide CO2 measurement range actually measured by infrared detector - not calculated
- NOx can be NO only or combined NO and NO2
- Rapid reading recovery on CO, NOx, or SO2 'overdose'
- Use on any fuel
- Pays for itself in months through fuel savings
Thông số kỹ thuật
- Method of Detection : Customer replaceable electrochemical O2, CO, SO2, and NO sensors; Solid state infrared detector for CO2
- Ranges Available : 0-1800°F (0-1000 °C) net stack temperature; 0-20.00% CO2; 0-200 or 0-2000 PPM NOx (as NO or NO + NO2); 0-200 or 0-2000 PPM SO2; 0-25.0 % O2; 0-9999 PPM CO
- Resolution : ±0.1% CO2, 1 PPM, 1°F (1°C)
- Accuracy & Repeatability : ±10 PPM for all gases; 0.1%
- Drift : <2% full scale per 8 hours of continuous operation
- Response Time (T-90) : < 30 seconds for 90% of O2, CO2, CO and SO2, < 60 seconds for 90% NO
- Ambient Temperature Range : Operating : 32 to 122°F (0 to 50°C); Sample : Up to 1200°F (635°C) (standard probe)
- Linearity : ±1% of full scale for each gas measured
- Power : 12 V battery operation, 115 VAC, 60 Hz for recharging; 220 VAC, 50 Hz
- Output Options : 0-1 V for each gas measured. RS232 or RS485, printer; 4-20 mA
- Analyzer Type : Flue Gas Analyzer