- The Nova 322 Portable Paramagnetic Analyzer is designed to measure the oxygen concetration of clean, dry sample gas. A built-in pressure regulator reduces any positive sample pressure. A built-in sample pump may be switched on to draw in the sample gas when it is at or slightly below atmospheric pressure. An external gas scrubber is available if required, to remove any corrosive gases prior to entering the analyzer.
- In operation, the incoming sample gas passes through sample conditioning filters and the flow meter, then into the O<sub>2</sub> sensor. The paramagnetic sensor utillizes a magneto-dynamic measuring cell and powerful selenium cobalt magnet assembly combined with on-board amplification, temperature compensation and barometric pressure compensation. This cell makes use of the principle that O<sub>2</sub> is drawn into a magnetic field, thereby increasing the turning force on a diamagnetic body suspended in the field. Very few other gases respond this way, making this method specific to O<sub>2</sub>. The Model 322 can be used for any O<sub>2</sub> measurement between 0-100 as long as the sample gas is clean, dry and non-corrisive.
- High accuracy paramagnetic O2 cell, very stable
- Fast response (T90-4 to 5 seconds)
- Bright digital readout
- Linear over entire 0-100% range
- Non-consumable O2 cell, no periodic replacement required
- Built-in sample pump or pressure regulator
- To improve accuracy, cell is compensated for changes in barometric pressure
- Bench Top (BT) enclosure is standard
Thông số kỹ thuật
- Method of Detection : Magneto-dynamic paramagnetic O2 cell
- Ranges Available : Any range from 0-2.0% to 0-100.0% O2
- Resolution : 0.1% O2
- Accuracy & Repeatability : Model 322BT: ± 1% of full scale; Model 322HABT: ± 0.5% of full scale
- Drift : ± 0.5% of full scale per month
- Response Time (T-90) : 4-6 seconds
- Ambient Temperature Range : 0-50 ºC; 32-122 ºF
- Linearity : Model 322BT: ± 1% of full scale; Model 322HABT: ± 0.5% of full scale
- Size & Weight : 8" W x 9" H x 10" D @ 12 lbs (20 x 23 x 25 cm @ 5.5kg)
- Power : AC/DC Operation, 115 VAC 60 Hz for recharging (Other voltages available)
- Output Options : 4-20mA or 0-1V (optional)
- Alarms : One high or one low O2 alarm (optional)
- Hazardous Area Rating Options : General Purpose, Non-Hazardous