Measurement Range
When using the CABLE-CO2 and a U12 or ZW: 0 to 2500 ppm
When using the CABLE-2070 and a H22 or U30: 0 to 4000 ppm
Telaire 7001 operating range: 32°F to 122°F (0°C to 50°C), 0 to 95% RH, non-condensing
Display resolution: ±1 ppm
Accuracy: ±50 ppm or 5% of reading, whichever is greater
Repeatability: ±20 ppm
Temperature dependence: ±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced at 25°C.
Pressure dependence: 0.13% of reading per mmHg (corrected via user input for elevation)
Response time: <60 seconds for 90% of step change
Warm-up time: <60 seconds at 72°F (22°C)
Calibration interval: 12 months Full factory calibration available by contacting GE Measurement & Control Solutions.
http://www.gesensinginspection.com/en/services-and-support.html
Battery type: Four AA batteries (not included)
Battery operation: 80 hours (alkaline)