TSI's Flowmeter design team can develop the precise mechanical configuration, performance characteristics, and electronic interface you need to optimize your design. The exceptional performance of these embedded flow sensors can enhance your product's performance and improve the time to market on new product development.
Temperature range: 0 to 50 °C
Pressure drop: 12 cm H2O at 300 l/min
Response time: 5 msec (0 to 63% of full-scale flow)
Output signals flow: 0 to 2.5 vdc non-linear, 2 channels
Output signal temperature: 0 to 2.5 vdc non-linear
Input voltage: 5 vdc ±5%, regulated, 3.3 vdc eeprom
Calibration: air
Flow range: 0 to 300 standard l/min
Accuracy*: ±2.5% of reading ±0.05 l/min @ 21°C (*Consult TSI for complete product specifications. Specifications are subject to change without notice.)
Power consumption: 0.8 watts at full-scale flow
Mating connector: Molex 51110-1060 (Wire) or Molex 87568-1074/87568-1073 (IDC)
Eeprom interface: I2C communication