General information |
|
Article number |
3.510.006.015 |
Particle concentration |
|
Number of channels |
6 |
|
Duct sizes |
0.3 µm, 0.5 µm, 1.0 µm, 2.5 µm, 5.0 µm, 10.0 µm |
|
Counting modes |
Concentration, cumulative, differential |
|
Counting efficiency |
50 % at 0.3 µm, 100 % for particles > 0.45 µm |
|
Flow rate |
2.83 l/min (0.1 ft³/min), controlled by internal pump |
|
Zero count |
< 1 particle / 5 min |
|
Coincidence loss |
5 %, 2 million particles per ft³ (28.3 litres) |
|
Light source |
laser class 3B, wavelength 780 nm, 90 mW |
|
Sample inlet |
Isokinetic probe |
Air temperature |
|
Min. measuring range [°C] |
0 |
|
Max. measuring range [°C] |
50 |
|
Min. measuring range [°F] |
32 |
|
Max. measuring range [°F] |
122 |
|
Accuracy |
10 – 40 °C ±0.5 °C
<10 °C ±1 °C
>40 °C ±1 °C
50 – 104 °F ±0.9 °F
<50 °F ±1.8 °F
>104 °F ±1.8 °F
|
Relative humidity |
|
Min. measuring range [%] |
0 |
|
Max. measuring range [%] |
100 |
|
Accuracy 40 % - 60 % [%] |
3 |
|
Accuracy 20 % - 40 % [%] |
3.5 |
|
Accuracy 60 % - 80 % [%] |
3.5 |
|
Accuracy 0 % - 20 % [%] |
5 |
|
Accuracy 80 % - 100 % [%] |
5 |
Dew point temperature |
|
Min. measuring range [°C] |
0 |
|
Max. measuring range [°C] |
50 |
|
Min. measuring range [°F] |
32 |
|
Max. measuring range [°F] |
122 |
|
Accuracy |
10 – 40 °C ±0.5 °C
<10 °C ±1 °C
>40 °C ±1 °C
50 – 104 °F ±0.9 °F
<50 °F ±1.8 °F
>104 °F ±1.8 °F
|
Wet-bulb temperature |
|
Min. measuring range [°C] |
0 |
|
Max. measuring range [°C] |
80 |
|
Min. measuring range [°F] |
32 |
|
Max. measuring range [°F] |
176 |
|
Accuracy ± [°C] |
1 |
|
Accuracy ± [°F] |
1.8 |
Interfaces |
|
USB |
|
|
3,5 headphone jack |
|
Memory organisation |
|
Internal flash memory |
Up to 5,000 data records |
|
Extended data memory (microSD, optional) - (max. 8 GB) |
|
|
Photo storage format |
JPEG (640 x 480) |
|
Video storage format |
3GP (320 x 240) |
|