- Primary standard for defining the pressure scale in a range up to 100 bar pneumatic
- Reference instrument for calibration laboratories and factory for the testing, adjustment and calibration of pressure measuring instruments
- Self-contained, complete, stand-alone system, also suitable for on-site use
- Total measurement uncertainty down to 0.008% of measured value
- Factory calibration includes traceability to national standars, as standard. Optional with DKD/DAkkS calibration
- High long-term stability with recommended recalibration cycle every 5 years
- Masses manufactured from stainless steel and aluminium, local gravity adjustment possible at no additional charge
- The optional quck-change system of piston-cylinder system enables fast and secure exchange of the piston-cylinder system in order to change the measuring range without tools.