Type |
LCC 200-1 |
LCC 200-10 |
LCC 200-100 |
LCC 200-1000 |
Order-Code |
LCC-200-1 |
LCC-200-10 |
LCC-200-100 |
LCC-200-1000 |
Pressure range |
-1...+1 mbar |
-10...+10 mbar |
-100...+100 mbar |
-1000...+1000 mbar |
Overpressure |
For delta p measurements max. system pressure 2 bar |
Measurement Uncertainty (+1 digit) |
±0.15% FS |
±0.05% FS |
±0.05% FS |
±0.05% FS |
Linearity ±1digit |
±0.1% FS |
±0.05% FS |
±0.05% FS |
±0.05% FS |
Hysteresis |
max. 0.05% FS |
Measurement principle |
inductive |
Temperature drift of the
internal reference sensor |
Zero: 0.03% FS / K (0% via zero balance)
Span: 0.03% FS / K |
Zero balance |
• automatic (in settable time spans, possible to swith of), or
• manuel (by pressing the ZERO button) |
Long-term stability of the
internal reference sensor |
0.1% FS per year (typical) |
Temperature ranges |
Working range +10...+40°C; Storage -10...+70°C |
Usable pressure measuring range |
-10 ... + 110% |
Adjustment accuracy |
0.01% FS |
Adjustment time |
depending on volume, <5 sec |
Measurement media |
Air, non-aggressive and corrosion-free dry gases |
Operation modes |
• CTRL (controlling pressure)
• MESS (measuring pressure)
• Remote-controlled operation (via interface) |
Display |
TFT touch screen, 7", colour |
Interface |
USB |
Supply |
24 VDC / 1 A, build-in Li-Ion battery (8h typical) |
Pressure ports |
6.6 x 11 mm (for flexible hose with 6 mm inner diameter) |
Dimension |
H 191,6 x W 257 x D 271 mm (without handel) |
Operating (Menu) languages |
German, English |