The 1090 is a precision portable process calibrator that combines source and measurement functions for thermocouples, PT100s, µV, mV & mA.
As a multifunction instrument the 1090 combines accuracy and durability with simple operation, making it ideal for process plant applications.
Thermocouple Measurement and Simulation
Measure and simulate the temperature and mV characteristics of J, K, T, R, S, B, N and E thermocouples.
Cold Junction Compensation
The unit can be operated with or without internal cold junction compensation.
PT100 Measurement and Simulation
Based on 0.3850 alpha probe standard. Range is -200°C to 700°C.
Measurement and Source (µV, mV, and mA)
Measurement ranges are 0 to +/-30mV and 0 to +/-60mA. Source ranges are 0 to +/-80mV and 0 to +80mA.
Temperature Units Selection
The display can be easily changed from °C to °F. The equivalent uV (thermocouples) and ohms (PT100) can also be shown.
24V Process Loop Drive Mode
A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA (or a lower level if required).
Inching (Incrementing/Decrementing)
The unit has a general-purpose inching function. This adjusts the output in fixed increments of temperature (thermocouples only) or voltage or current. The set-up menu gives a the user a choice of three levels of increment i.e. 0.1, 1 or 10 for °C/°F, or 1, 10, or 100 uV/uA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. This is especially useful for calibrating thermostat controllers that have tight specification on hysteresis.
|
Thermocouple Type |
Temperature Range (°C) |
Accuracy (°C) |
J
|
-210 to 150
150 to 1200
|
0.15
0.3
|
K
|
-270 to 190
190 to 1250
|
0.5
0.4
|
T
|
-200 to 150
150 to 400
|
0.4
0.5
|
R
|
-50 to 800
800 to 1750
|
0.8
2.0
|
S
|
-50 to 850
850 to 1750
|
0.9
2.0
|
B
|
100 to 1200
1200 to 1800
|
2.0
3.0
|
N
|
-270 to 260
260 to 1300
|
0.5
1.0
|
E |
-50 to 1000 |
0.3 |
|
|
|
Resolution
|
0.1°C or 0.1°F
|
|
|
Thermocouple Type |
Temperature Range (°C) |
Accuracy (°C) |
J |
-200 to 580 |
0.7 |
K
|
-200 to -150
-150 to 750
|
2.5
0.5
|
T
|
-200 to 0
0 to 400
|
1.5
0.4
|
R
|
50 to 400
400 to 1750
|
3.0
1.5
|
S
|
-50 to 100
100 to 1750
|
3.0
1.5
|
B
|
110 to 1000
1000 to 1800
|
3.5
1.5
|
N |
-100 to 890 |
0.6 |
E |
-50 to 400 |
0.4 |
|
|
|
Resolution
|
0.1°C or 0.1°F
|
|