AC withstand voltage test |
Rated output capacity |
200VA (5000V/40mA), short circuit current is greater than 200mA;
(Optional: 500VA (5000V/100mA), short-circuit current greater than 200mA;) |
Output voltage |
Range/accuracy |
100~5000V, ±(1%× setting +5V) |
Output frequency |
Range/accuracy |
50Hz or 60Hz, ±0.1%× setting |
Alarm current |
Upper limit/accuracy |
0.10~40.00mA, ±(1%× setting +5 words) |
Lower limit /accuracy |
0.000~9.999mA, ±(1%× setting +5 words) |
Time |
Range |
0.5~999.9s (0: infinite time) |
Ramp-up time |
0.1~999.9s (0: off) |
Ramp-down time |
0.1~999.9s (0: off) |
Accuracy |
± (0.2%× setting +1 word) |
Measure meter |
Voltage range/accuracy |
0.01~5.00kV, ± (1%× reading +2 words) |
Current range/accuracy |
0.10~40.00mA, ± (1%× reading +5 words) |
Time range/accuracy |
0.1~999.9s, ± (0.2%× reading +1 word) |
Output voltage wave, distortion, adjusting rate |
Sine wave, less than 2% (pure resistive load), ± (1% × setting + 5V) (from zero to full load) |
Start voltage |
(0%~50%)× output voltage setting |
Current compensation |
0.000~40.00mA, Auto/manual |
Arc detection |
Level 1~9, 9: highest level; 0: off |
DC withstand voltage test |
Output voltage |
Range/accuracy |
100~6000Vdc, ± (1%× setting +5V) |
Alarm current |
Upper limit/accuracy |
0~10000μA, ± (1%× setting +5 words) |
Lower limit /accuracy |
0.0~999.9μA, ± (1%× setting +5 words) |
Time |
Range |
0.5~999.9s(0: infinite time) |
Ramp-up time |
0, 0.4~999.9s (0: off) |
Ramp-down time |
0, 1.0~999.9s (0: off) |
Accuracy |
± (0.2%× setting +1 word) |
Measure meter |
Voltage range/accuracy |
0.01~6.00Vdc,±(1%×读数值+2个字)
0.01~6.00Vdc, ± (1%× reading +2 words) |
Current range/accuracy |
0.0~10000uA, ± (1%× reading +5 words) |
Time range/accuracy |
0.1~999.9s, ± (0.2%× reading +1 word) |
Rated output capacity |
60VA(6000Vdc/10mA) |
Output adjustment |
± (1%× setting +2V), zero to full load |
Compensation current |
0 ~200.0μA, Auto/manual |
Start voltage |
(0%~50%)×output voltage setting |
Charging current lower limit |
(0~3.500) μA, Auto/manual |
Discharge time |
≤200ms |
Arc detection |
Level 1~9, 9: highest level; 0: off |
Insulation resistance test |
Output voltage |
Range/accuracy |
100~2500Vdc, ± (1%× setting +5V) |
Alarm resistance |
Upper/lower limit |
0.1MΩ~999.99MΩ,1000MΩ~50000MΩ,上限包含无上限设定
0.1MΩ~999.99MΩ, 1000MΩ~50000MΩ, the upper limit includes no upper limit setting |
Accuracy |
100-499V, 0.10MΩ~999.99MΩ, 1000~2000MΩ, ± (5%× setting +2 words)
500-2500V, 0.10~999.99MΩ, ± (2%× setting +2 words),
1000~9999MΩ, ± (5%× setting +2 words)
10000~50000MΩ, ± (15%× setting +2 words) |
Time |
Ramp-up time |
0,0.1~999.9s (0: off) |
Delay decision time |
0, 0.5~999.9s (0: infinite) |
Range |
0, 0.5~999.9s (0: infinite) |
Ramp-down time |
0, 1.0~999.9s (0: off) |
Accuracy |
± (0.1%× setting +0.1s) |
Measure meter |
Voltage range/accuracy |
100~2500Vdc, ± (2%× reading +5 words) |
Resistance range/accuracy
|
Range: 0.100 MΩ~9.999 MΩ, 10.00 MΩ~99.99 MΩ, 100.0 MΩ~999.9 MΩ, 1.000 GΩ~50.00GΩ
Resolution: 0.001 MΩ, 0.01 MΩ, 0.1 MΩ, 0.001GΩ, 0.01GΩ
100-499V, 0.10~9.999MΩ, 10.00~99.99MΩ, 100.0~999.9MΩ, 1~2GΩ, ± (5%× reading +2 words)
500-2500V, 0.100~9.999MΩ, 10.00~99.99MΩ, 100.0~999.9MΩ, ± (2%× reading +2 words),
1.000~9.999GΩ, ± (5%× reading +2 words)
10.00~50.00GΩ, ± (15%× reading +2 words) |
Time range/accuracy |
0.1~999.9s, ± (0.2%× reading +1 word) |
Charging current lower limit |
0~3.500μA, Auto/manual |