AC Hipot test
|
Rated Output
|
200VA(5000V/40mA),short circuit current is greater than 200mA;
|
Output Voltage
|
Range/accuracy
|
100~5000V,±(1%of setting+2 counts)
|
Output Frequency
|
Range/accuracy
|
50Hz/60Hz,±0.1% of setting
|
Alarm Current
|
Upper limit/ accuracy
|
0.10~40.00mA,±(1%of setting+2 counts)
|
Lower limit/ accuracy
|
0.000~9.999mA,±(1%of setting+2 counts)
|
Start Voltage
|
(0%~50%)× output voltage setting
|
Current compensation
|
0.000~40.00mA,Auto/Manual
|
ARC detection
|
1~9 level,9level is the most senstive,0 is off
|
DC Hipot test
|
Output Voltage
|
Range/accuracy
|
100~6000Vdc,±(1%of setting+2 counts)
|
Alarm Current
|
Upper limit/ accuracy
|
0~10000μA,±(1%of setting+2 counts)
|
Lower limit/ accuracy
|
0.0~999.9μA,±(1%of setting+2 counts)
|
Current compensation
|
0 ~200.0μA,Auto/Manual
|
Start Voltage
|
(0%~50%)×output voltage setting
|
Charging Current lower Limit
|
(0~3.500)μA,Auto/Manual
|
Discharge Time
|
≤200ms
|
ARC detection
|
1~9 level,9 level is the most senstive,0 is off
|
Insulation Resistance test
|
Output Voltage
|
Range/accuracy
|
100~2500Vdc,±(1%of setting+2V)
|
Alarm Current
|
Upper limit/ accuracy
|
100-499V,1~1000MΩ,±(8%of setting+2 counts)
1000~2000MΩ,±(12%of setting+2 counts)
500-2500V,1~199MΩ,±(2%of setting+2 counts),
200~999MΩ, ± (5%of setting+2 counts)
1000~9999MΩ, ± (15%of setting+2 counts)
|
Lower limit/ accuracy
|
Ground Bond test
|
Rated Output
|
Max current 32A,Max test resistance:600mΩ,max open circuit voltage:12V
|
Output Current
|
Range/accuracy
|
2.0~32.0A,±(1%of setting+2 counts)
|
Output Voltage
|
Range/accuracy
|
3.0~6.4V ,±(1%of setting+2 counts)
|
Output Frequency
|
Range/accuracy
|
Sin wave,50/60Hz,±0.1% of setting
|
Alarm Limit
|
Resistance upper/lower limit |
2.0~10.6A:0.1~600mΩ;
|
10.7~32.0A:0.1~R mΩ, R=(6400 / current setting)mΩ,As the current range is switched, the tester will auto calculate the max range
|
Accuracy
|
±(1%of setting+2 counts)
|
Compensation Resistance
|
Range
|
0~100 mΩ,automatic measurement, compensation ON/OFF
|