AC Hipot test
|
Rated output capacity |
AN9638H(F) |
500VA(5000V/100mA),short circuit current >200mA |
AN9637H(F) |
200VA(5000V/40mA),short circuit current >200mA |
Output voltage setting |
Range,Accuracy |
100~5000V,±(2%×setting+5counts) |
Output frequency setting |
Range,Accuracy |
50Hz or 60Hz,±0.1%×setting |
Alarm current limits setting |
Upper-limit range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×setting+5counts) |
Lower-limit range,Accuracy |
0.000~9.999mA,±(2%×setting+5counts) |
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up/down time range |
0.1~999.9s |
Accuracy |
±(0.2%×setting+1count) |
Measurement |
Voltage range,Accuracy |
0.01~5.00kV,±(1.5%×reading+1count) |
Frequency range,Accuracy |
50Hz or 60Hz,±0.1%×reading |
Current range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×reading+2counts) |
Time range,Accuracy |
0.1~999.9s,±(0.2%×reading+1count) |
Output voltage waveform, distortion, regulation |
Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load) |
Start voltage setting |
(0%~50%)×output voltage setting |
Current compensation setting |
0.000~100.00mA(or 40.00mA),auto、manual |
Arc detection |
1~9 level,9 level is optimally sensitive,0 is OFF |
DC Hipot test
|
Output voltage setting |
Range,Accuracy |
100~6000VDC,±(2%×setting+5counts) |
Alarm current limits setting |
Upper-limit,Accuracy |
0~10000μA,±(2%×setting+5counts) |
lower-limit,Accuracy |
0.0~999.9μA,±(2%×setting+5counts) |
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up time range |
0.1~999.9s |
Ramp down time range |
0.3~999.9s(0 is OFF) |
Accuracy |
±(0.2%×setting+1 count) |
Measurement |
Voltage range,Accuracy |
0.01~6.00VDC,±(1.5%×reading+1count) |
Frequency range,Accuracy |
0.0~10000uA,±(2%×reading+5count) |
Current range,Accuracy |
0.1~999.9s,±(0.2%×reading+1count) |
Rated output capacity |
60VA(6000VDC/10mA) |
Output voltage ripple, regulation |
<5%(pure resistance load,6000V/10mA),<(2%×setting+5V)(from empty load to full load) |
Compensation current Setting |
0 ~200.0μA,auto, manual |
Start voltage setting |
(0%~50%)× setting output voltage |
Charging current lower limit |
(0~3.500)μA,auto, manual |
Discharge time |
≤200ms |
Arc detection |
1~9 level,9 level is optimally sensitive,0 is OFF |