AC Withstand Voltage Test
|
Rated output
|
200VA (5000V/40mA), short circuit current is greater than 200mA;
[Optional: 5000V/100mA]
|
Output voltage
|
Range/accuracy
|
100~5000V, ± (2%× setting + 5 words)
|
Output frequency
|
Range/accuracy
|
50Hz or 60Hz, ±0.1%× setting
|
Alarm current
|
Upper limit/accuracy
|
0.10~ or 40.00mA, ± (2%× setting + 5 words)
|
Lower limit /accuracy
|
0.000~9.999mA, ± (2%× setting + 5 words)
|
Time
|
Range
|
0.5~999.9s (0: infinite time)
|
Ramp time
|
0.1~999.9s
|
Descent time
|
0.1~999.9s
|
Accuracy
|
± (0.2%× setting + 1 word)
|
Measure meter
|
Voltage range/accuracy
|
0.01~5.00kV, ± (1.5%× reading + 1 word)
|
Frequency range/accuracy
|
50Hz or 60Hz, ±0.1%× reading
|
Current range/accuracy
|
0.10~40.00mA, ± (2%× reading + 2 words)
|
Time range/accuracy
|
0.1~999.9s, ± (0.2%× reading + 1 word)
|
Output voltage wave, distortion, adjusting rate
|
Sine wave, less than 2% (pure resistive load), less than (2% × setting + 5V) (from zero to full load)
|
Start voltage
|
(0%~50%)× output voltage setting
|
Current compensation
|
0.000~40.00mA, Auto/manual
|
Arc detection
|
Level 1~9, 9: highest level; 0: off
|
DC withstand voltage test
|
Output voltage
|
Range/accuracy
|
100~6000Vdc, ± (2%× setting + 5 words)
|
Alarm current
|
Upper limit/accuracy
|
0~10000μA, ± (2%× setting + 5 words)
|
Lower limit /accuracy
|
0.0~999.9μA, ± (2%× setting + 5 words)
|
Time
|
Range
|
0.5~999.9s (0: infinite time)
|
Ramp time
|
0.1~999.9s
|
Descent time
|
0.3~999.9s (0, off)
|
Accuracy
|
± (0.2%× setting + 1 word)
|
Measure meter
|
Voltage range/accuracy
|
0.01~6.00Vdc, ± (1.5%× reading + 1 word)
|
Current range/accuracy
|
0.0~10000uA, ± (2%× reading + 5 words)
|
Time range/accuracy
|
0.1~999.9s, ± (0.2%× reading + 1 word)
|
Rated output
|
60VA (6000Vdc/10mA)
|
Output voltage fluctuation, regulation
|
Less than 5% (purely resistive load, 6000V/10mA), less than (2% × setting + 5V) (from zero to full load)
|