The V-750 UV-Visible Spectrophotometer optical design uses precision double-beam optics with variable spectral bandwidth. A high sensitivity PMT detector provides accurate and reproducible measurements for low to high concentration samples. By controlling the high voltage applied to the PMT, the dynode feedback offers a wider dynamic range than is generally found on uv-visible spectrophotometers.

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Optical System Czerny-Turner mount

Single monochromator

Fully symmetrical double beam type
Light Source Halogen lamp, Deuterium lamp
Wavelength range 190 to 900 nm
Wavelength accuracy +/-0.2 nm (at 656.1 nm)
Wavelength repeatability +/-0.05 nm
Spectral bandwidth (SBW) 0.1, 0.2, 0.5, 1, 2, 5, 10 nm

L2, L5, L10 nm (low stray light mode)

M1, M2 nm (micro cell mode)
Stray light 1 % (198 nm KCL 12 g/L aqueous solution)

0.005 % (220 nm NaI 10 g/L aqueous solution)

0.005 % (340 nm NaNO2 50 g/L aqueous solution)

0.005 % (370 nm NaNO2 50 g/L aqueous solution)

SBW: L2 nm
Photometric range -4~4 Abs
Photometric accuracy +/-0.0015 Abs (0 to 0.5 Abs)

+/-0.0025 Abs (0.5 to 1 Abs)

+/-0.3 %T 

Tested with NIST SRM 930D
Photometric repeatability +/-0.0005 Abs (0 to 0.5 Abs)

+/-0.0005 Abs (0.5 to 1 Abs)

Tested with NIST SRM 930D
Scanning speed 10~4000 nm/min (8000 nm/min in preview mode)
Slew speed 12,000 nm/min
RMS noise 0.00003 Abs 

(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm)
Baseline stability 0.0003 Abs/hour

(Wavelength: 250 nm, response: slow and SBW: 2 nm)
Baseline flatness +/-0.0002 Abs (200 - 850 nm)
Detector Photomultiplier tube
Standard functions IQ Accessory, Start Button, Analog output
Standard programs Abs/%T meter, Quantitative analysis, Spectrum measurement, 

Time course measurement, Fixed wavelength measurement, Validation, 

Daily Check 

Two wavelength time course measurement
Dimensions and weight 460(W) x 602(D) x 268(H) mm, 27 kg
Power requirements 150 VA
Installation requirements Room temperature: 15-30 Celsius, humidity: below 85%