Optical System |
Czerny-Turner mount
Single monochromator
Fully symmetrical double beam type |
Light Source |
Halogen lamp, Deuterium lamp |
Wavelength range |
190 to 1600 nm |
Wavelength accuracy |
+/-0.3 nm (at 656.1 nm)
+/-1.0 nm (at 1312.2 nm) |
Wavelength repeatability |
+/-0.05 nm (UV-Vis), +/-0.1 nm (NIR) |
Spectral bandwidth (SBW) |
UV-Vis:
0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray light mode)
M1, M2 nm (micro cell mode)
NIR:
0.2, 0.4, 0.5, 1, 2, 4, 10, 20,
L4, L10, L20 nm (low stray light mode)
M2, M4 nm (micro cell mode) |
Stray light |
1 % (198 nm KCL 12 g/L aqueous solution)
0.005 % (220 nm NaI 10 g/L aqueous solution
0.005 % (340 nm NaNO2 50 g/L aqueous solution)
0.005 % (370 nm NaNO2 50 g/L aqueous solution)
SBW: L2nm
0.04 % (1420 nm: H2O)
SBW: L4 nm |
Photometric range |
UV-Vis: -4~4 Abs
NIR: -3~3 Abs |
Photometric accuracy |
+/-0.0015 Abs (0 to 0.5 Abs)
+/-0.0025 Abs (0.5 to 1 Abs)
+/-0.3 %T Tested
with NIST SRM 930D |
Photometric repeatability |
+/-0.0005 Abs (0 to 0.5 Abs)
+/-0.0005 Abs (0.5 to 1 Abs)
Tested with NIST SRM 930D |
Scanning speed |
10~4000 nm/min (8000 nm/min in preview mode) |
Slew speed |
UV-Vis: 12000 nm/min
NIR: 24000 nm/min |
RMS noise |
0.00003 Abs
(0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW:2 nm) |
Baseline stability |
0.0003 Abs/hour
0.04 % (220 nm NaI 10 g/L aqueous solution)
0.02 % (340 nm NaNO2 50 g/L aqueous solution)
0.02 % (370 nm NaNO2 50 g/L aqueous solution)
(10 mm cell) |
Baseline Stability |
±0.0004 Abs/hour
(Wavelength: 250 nm,
response: slow and SBW: 2nm.) |
Baseline flatness |
+/-0.0002 Abs (200 - 1600 nm) |
Detector |
Photomultiplier tube, Peltier cooled InGaAs photodiode |
Standard functions |
IQ Accessory, Start Button, Analog output |
Standard programs |
Abs/%T meter, Quantitative analysis, Spectrum measurement,
Time course measurement, Fixed wavelength measurement, Validation,
Daily Check
Two wavelength time course measurement |
Dimensions and weight |
460(W) x 602(D) x 268(H) mm, 29 kg |
Power requirements |
150 VA |
Installation requirements |
Room temperature: 15-30 Celsius, humidity: below 85% |