FEATURES |
|
■ |
Wide measurable wow flutter range 0.0015% for testing the highest grade to general-purpose recording and playback equipment. |
■ |
Accurate measurement under different standards: Effective value for RMS, average value for NAB, and peak-value for IEC (DIN, CCIR). |
■ |
Wow flutter measurements under standard weighted and specified unweighted conditions; in addition, wow and flutter values can be separately determined. |
■ |
Monitor lamp indicates the proper input voltage level for wow flutter and tape speed measurements. |
■ |
Stable and accurate signal output at 3kHz and 3.15kHz (crystal-controlled) available for recording a test tape, disc, etc. |
■ |
FIM measurements possible in accordance with DIN45411 specification. |
Input Frequencies |
3kHz ±300Hz, for RMS, NAB. |
3.15kHz ±300Hz, for IEC (DIN, CCIR). |
|
Input Voltage Range |
Two range: 0.1 to 30mVrms, and 5mV to 10Vrms
|
|
Input Impedance |
Approx. 300kΩ unbalanced |
|
Wow Flutter Range |
0.0015% to 3% in six ranges
|
0.003% to 3% at 0.1 to 30mV input |
0.01%, 0.03%, 0.1%, 0.3%, 1% and 3% at full scale
|
|
Meter Indication |
Peak value for IEC (DIN, CCIR) |
Average value for NAB |
Effective value for RMS
|
|
Indication Accuracy |
±5% of full scale at 4Hz |
|
Frequency Characteristics |
Weighted |
0.2 to 200Hz in accordance with standards for IEC (DIN, CCIR)
and NAB. |
Wow Flutter separation |
• Wow |
0.5Hz to 6Hz |
• Flutter |
6Hz to 200Hz |
Unweighted |
• NAB |
0.5 to 200Hz -3dB±1dB |
• IEC(DIN, CCIR) |
0.3 to 200Hz -3dB±1dB |
Unweighted |
Standards |
Response |
Roll off |
IEC (DIN, CCIR) |
0.3 to 200Hz
-3dB±1dB |
<-6dB/oct below 0.3Hz
<-15dB/oct above 200Hz |
NAB |
0.5 to 200Hz
-3dB±1dB |
<-6dB/oct below 0.5Hz
<-15dB/oct above 200Hz |
|
|
|
FIM Measurement |
Range |
Corresponding to Wow Flutter |
Filter Characteristics |
• FIM : ON |
L.P.F. |
500Hz |
-3dB±1dB |
Roll off |
above 500Hz |
<-36dB/oct |
|
|
• 160HzFIL. : ON |
H.P.F. |
100Hz |
-3dB±1dB |
Roll off |
above 100Hz |
<-12dB/oct |
|
|
|
Tape Speed Range |
3kHz ±1kHz : 4-digit indication |
|
Internal Recording signal |
Frequency |
3kHz and 3.15kHz ±3×10-5 (crystal-controlled) |
Output Voltage |
Approx. 0.2Vrms at open circuit |
Output Impedance |
Approx. 600Ω unbalanced |
Distortion |
Less then 0.5% |
|
Drift |
Output voltage |
±1V per ±10% drift |
Output impedance |
600Ω ±10% unbalanced |
|
To Recorder |
Output voltage |
DC 1V±5% at full scale |
Output impedance |
600Ω ±10% unbalanced |
|
Flutter |
Output voltage |
1Vrms±5% at full scale |
Output impedance |
600Ω ±10% unbalanced |
|
Frequency Counter (at independent operation) |
Range |
10Hz to 99.99kHz in two ranges |
0.01 to 9.999kHz, 0.1 to 99.99kHz |
Accuracy |
±(1 count+reference frequency accuracy) |
Reference Frequency |
600kHz ±1×10-4 (0 to 40ºC) |
Input Impedance |
Approx. 300kΩ unbalanced |
Gate Time |
0.1 sec, 1 sec in 2 ranges |
Indication |
4-digit display |
|
General Data |
Power Requirements |
AC 100V, 115V, 215V, or 230V±10%
50/60Hz;
approx. 15VA
|
Dimensions, overall |
Approx. 360(W)×165(H)×290(D)mm |
Weight |
Approx. 5kg |
Accessory, furnished |
Input/Output cord : 1ea |
|
* |