Compensated Flow Range
|
1000 to 3000 ml/min, ± 5%
(5 to 500 ml/min requires low flow accessories; see Low Flow Kit)
Requires 1/4-inch ID tubing
|
Flow Control
|
Patented (U.S. Patent No. 6,741,056) system holds constant flow to ± 5% of the set-point |
Compensation Range for a Minimum of 8-hour Operation
|
1000 ml/min at 30 inches water back pressure
2000 ml/min at 20 inches water back pressure
3000 ml/min at 10 inches water back pressure |
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min) |
1.0 |
1.5 |
2.0 |
2.5 |
3.0 |
Filter/Pore Size (µm) |
|
|
|
|
|
25-mm MCE/0.8 |
6 |
9 |
12 |
15 |
— |
25-mm MCE/0.45 |
14 |
22 |
— |
— |
— |
37-mm MCE/0.8 |
2 |
3 |
4 |
5 |
6 |
37-mm PVC/5.0 |
1 |
1 |
2 |
2 |
2.5 |
Compare the information in this table to pump compensation range to determine appropriate applications.
|
Temperature Range
|
Operating: 32 to 104 F (0 to 40 C)
Storage: -4 to 113 F (-20 to 45 C)
Protect sample pump from weather when in use outdoors.
|
Compensated Temperature Range
|
41 to 122 F (5 to 50 C ) |
Operating Humidity
|
0 to 95% non-condensing |
Power
|
Three standard AA alkaline batteries
Rechargeable AA-size 1.2-volt NiMH batteries may be used but will provide approximately half the run time stated for disposable batteries.
|
LED Functions
|
Slow flash = normal run
Fast flash = flow fault
Double flash = low battery, approximately one-hour run time remaining
Flash every eight seconds = pump shutdown due to exhausted batteries
Indication of pump shutdown due to exhausted batteries will override all other indicators.
Four flashes every six seconds = pump shutdown due to flow fault
|
Run Time
(new batteries)
|
8.5 hours at 1 L/min, 25 inches water back pressure
10 hours at 2.5 L/min, 15 inches water back pressure
Dependent on batteries used. See tables below.
|
Tubing |
Requires 1/4-inch ID tubing |
Case
|
Plastic, RFI/EMI-shielded, CE marked
|
Dimensions
|
4.6 x 3 x 1.75 in (11.7 x 7.6 x 4.4 cm) includes inlet and belt clip
|
Weight
(includes batteries)
|
12 oz (340 gm)
|