- Magnetic field technology of the "third generation"
+ The variable magnetic field strength up to 1.0 Tesla and the use of two different background correction modes guarantee maximum sensitivity and optimal adaptation to the application. The 3Field technology extends the linear working range and the dynamic mode ensures automatic adaptation to changing element contents.
- Transversely heated graphite furnace
+ The cross-heated graphite furnace offers optimal conditions for sample analysis, even in difficult matrices and in the measurement of refractory elements, by:
Homogeneous temperature distribution
Low atomization temperatures
Reduction of gas phase interference
Linear, fast heating rates
Minimization of memory effects
STPF concept
- The ZEEnit series includes the ZEEnit 650 P, a Zeeman AAS for graphite furnace and HydrEA technology, and the ZEEnit 700 P, a compact tandem spectrometer for flame, hydride, graphite furnace and hydrEA technology. Both devices are capable of analyzing both liquid and solid samples in the same device.