Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Bộ chuyển đổi nhiệt độ, HMT 334, Vaisala VietNam

Mã sản phẩm: HMT334
Sử dụng cho Model:
Hãng SX: Vaisala
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

The Vaisala HUMICAP® Humidity and Temperature Transmitter HMT334 is designed to measure humidity in pressurized spaces or vacuum chambers. Each probe is tested to ensure a gas and vacuum tight installation. 


Vaisala HUMICAP® performance 
The HMT334 incorporates Vaisala’s more than 30 years of experience in industrial humidity measurement. The sensor provides accurate and reliable measurement and is immune to particulate contamination and most chemicals. 

Graphical display of measurement trends and real time history 
The HMT330 series features a numerical and graphical display. The user can easily monitor operational data, measurement trends and up to one-year history. 

The optional data logger with realtime clock makes it possible to generate more than four years of measured history, and zoom in on any desired time or time frame. 

Using a serial line, the measurement data can be transferred to a PC where it can be further processed and copied to other programs.

▪ For pressures up to 100 bar and vacuum applications 
▪ For temperatures -70 ... +180 °C (-94 ... +356 °F) 
▪ ISO and NPT threads available 
▪ Application example: test chambers

Relative humidity
Measurement range 0 ... 100 % RH
Accuracy (including non-linearity, hysteresis, and repeatability) with Vaisala HUMICAP® 180 or 180R for typical applications
with Vaisala HUMICAP® 180C or 180RC for applications with chemical purge/warmed probe
at +15 ... +25 °C (59 ... +77 °F) ±1 %RH (0 ... 90 %)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH
at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 % RH (0 ... 40 %RH)
± 1.0 % RH (40 ... 97 %RH)
Response time (90 %) at +20 °C (+68 °F) 8 s / 17 s* with grid filter
in still air 20 s / 50 s* with grid + steel netting filter
40 s / 60 s* with sintered filter
Temperature
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.36 °F)
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 HZ
Power consumption at +20 °C (Uin 24 VDC) RS-232 max. 25 mA
Uout 2 x 0 ... 1 V/0 ... 5 V/0 ... 10 max. 25 mA
Iout 2 x 0 ... 20 mA max. 60 mA
display and backlight + 20 mA
during chemical purge max. 110 mA
during probe heating (HMT337) + 120 mA
Analog outputs (2 standard, 3rd optional) current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0... 1 V, 0... 5 V, 0... 10 V
Accuracy of analog outputs at +20 °C ±0.05% full scale
Temperature dependence of the
analog outputs
±0.005%/°C full scale
External loads current outputs RL < 500 ohm
0 ... 1 V output RL > 2 kohm
0 ... 5 V and 0 ... 10 V outputs RL > 10 kohm
Max. wire size 0.5 mm2 (AWG 20) stranded wires recommended
Digital outputs RS-232, RS-485 (optional)
Service connection RS-232, USB
Relay outputs (optional) 0.5 A, 250 VAC
Ethernet interface (optional) Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link
Ethernet interface (optional) Supported standards 802.11b
Connector Type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
Authentication / Encryption (WLAN) Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fixed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display of any parameter
Menu languages English, Chinese, Japanese, Spanish, German, French, Swedish, Russian, Finnish

Bình luận