Accuracy (at applied pressure):
Ranges < 125 Pa = ±2 Pa
Ranges 125 Pa = ±1,5 % Pa
(Accuracy specifications include: general accuracy,
temperature drift, linearity, hysteresis, long term
stability, and repetition error)
Thermal effects:
Temperature compensated across the full spectrum of
capability
Overpressure:
Proof pressure: 25 kPa
Zero point calibration:
Automatic autozero or manual pushbutton
Response time:
1.0−20 s, selectable via menu
Media compatibility:
Dry air or non-aggressive gases
Pressure units (select via menu):
Pa, kPa, mbar, inWC, mmWC