- Measures total combustible gases (TCG) in the field.
- Built-in gas pressure sensor.
- Includes regulators and accessories.
- Includes cylinders of Instrument Air, Zero Grade Nitrogen and 1 % Methane in Nitrogen calibration gases.
- All shipments containing gas cylinders ship from the factory via FedEx Ground ONLY. Reshipping this non-flammable calibration gas through common carrier requires hazardous materials certification because the cylinder pressure is greater than 40 psi.
Thông số kỹ thuật
Inlet Pressure |
1 PSIG to 5 PSIG, nominal
|
Damage Inlet |
Vacuum to 30 PSIG
|
Gas Inlet |
Female Luer Fitting
|
Damage Limit |
Vacuum to 30 PSIG
|
Gas Inlet |
Female Luer Fitting
|
Purity |
Zero-Grade Instrument Air, 21.00 % Oxygen
|
Inlet Pressure |
20 PSIG to 25 PSIG or use supplied regulator
|
Air Inlet |
Female Luer Fitting
|
Range |
0 % to 10.0 %
|
Accuracy |
± (5 % of reading + 1 count)
|
Resolution |
0.01 % TCG
|
Calibration |
1 % Methane = 1 % TCG
|