Features:
- 11 selectable pressure units
- O.05% full scale accuracy
- Pressure ranges 10 to 20,000 psi (1400 bar)
- Large, easy to read display with five digit resolution
- % pressure indication with bar graph for visual reference
- Temperature compensated accuracy from 14°F to 122°F (-10°C to 50°C)
- Pressure switch test, Min/Max, tare and alarm functions
- Adjustable mounting positions
- IDOS compatible with RS 232 interface and networking capability
- Stainless & Inconel wetted parts for aggressive media
- 0-5V analogue output
- Pump Kits available
Applications:
- Process monitoring and control
- Test and calibration
The DPI 104 can be used as a stand-alone process indicator or as a pressure reference for use with our widely popular hand pumps.
Specification:
Advanced silicon technology delivers 0.05% accuracy over ranges to 20,000 psi, min/max, tare and alarm functions, analog output and a pressure switch test function making the DPI 104 suitable for a wide array of pressure sensing applications.
Accuracy
|
0.05% FS including non-linearity, hysteresis,repeatability and temperature effects from 14°F to 122°F (-10°C to 50°C)
|
Burst Pressure
|
Burst pressure is 2x working pressure (exception, 20,000 psi (1400 bar); burst pressure 29,000 psi (2000 bar).
|
Selectable Pressure Units
|
kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O and mmHg
|
Burst Pressure
|
Burst pressure is 2x working pressure (exception, 20,000 psi (1400 bar); burst pressure 29,000 psi (2000 bar).
|
Selectable Pressure Units
|
kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O and mmHg
|
Display
|
Pressure reading: 5 digits with 0.5 in (12.7 mm) character height
Full scale indicator: 2.5 digits with 0.25 in (6.35 mm) character height
|
Pressure Bar Graph
|
In addition to the numeric pressure indication, the DPI 104 LCD display contains a circular twenty segment bar graph to provide the user with a quick visual estimation of pressure from 0 to 100% full scale output. The bar graph increments represent 5% of the user-selectable range.
|
Display Update
|
Two times per second
|
Minimum/Maximum Values
|
The minimum and maximum pressure values can be displayed on the DPI 104. This function can be enabled/disabled, and reset by the user.
|
Switch Test
|
The DPI 104 features a switch test function that will capture and display the open and closed values from an external pressure switch. Maximum switch impedance 200 O.
|
Voltage Output
|
The DPI 104 can be programmed to provide a 0 to 5 V output signal that can be configured as directly proportional to the pressure shown in the display or set to a fixed value. The voltage output mode provides 0.1% FS accuracy from 50 mV to 5 V.
|
Alarm Output
|
The alarm output consists of an open drain field effect transistor (maximum current 250 mA, maximum voltage 24 VDC)
|
Adjustable Mounting Positions
|
For added convenience, the DPI 104 housing will rotate 320° around the pressure fitting and the faceplate can be rotated in any orientation for optimum visibility.
|
Menu Lock
|
To guard against unauthorized menu access, the DPI 104 features a menu and tare lock function
|
RS232 Interface
|
Serial communications are provided to allow transfer of data to a PC with the optional serial lead (IA4090-2- V0). Using this link all menu commands and display data are available via an ASCII command set or the optional SiCal Pro Software.
|
|