Features:
- Ranges from 25 mbar to 700 bar
- Single or dual range configuration
- Pressure test and maintenance
- Transmitter calibration
- Loop set-up and diagnostics
- Switch verification
Applications:
- Pressure test and maintenance
- Transmitter calibration
- Loop set-up and diagnostics
- Switch verification
The DPI 800 Series is a complete range of advanced, robust and simple to use hand-held instruments. Highly cost effective, these tools are ideal for test/calibration of many popular process parameters. Advanced features and technical innovations address more applications in less time and deliver results you can rely on.
Specifications: (Contact us for fully detail)
Pressure Test and Measurement
DPI 800 Pressure Indicator
|
The ideal tool for pressure test and measurement
|
PressureRanges
|
10 in H2O to 10,000 psi (25 mbar to 700 bar) including vacuum options
|
All-Inclusive Accuracy
|
Can be relied on from one year to the next, even in tough environmental conditions (see specifications)
|
Dual Sensor Configuration
|
Extended measurement range and simultaneous two channel reading (P1 and P2 or P1 - P2)
|
Stainless Steel Sensor Construction
|
Available for compatibility with a wide range of fluids and gases (refer to range table)
|
Programmable Leak Test
|
Reports the pressure drop and leak rate
|
Advanced Features
|
Hold, maximum/minimum/average, alarm, and tare facilitate troubleshooting
|
Pressure Instrument and Loop Maintenance
DPI 802 Pressure Loop Calibrator
|
Provides simultaneous pressure and mA measurement for transmitter and loop maintenance
|
Dual Readings
|
Simultaneous measurement of pressure and mA for transmitter calibration and loop maintenance
|
24V Loop Power Supply
|
Energizes transmitters and control loops
|
Automatic Switch Test
|
Captures open/closed trip values providing a fast and highly accurate “safety system” check
|
HART Resistor
|
Can be switched into the loop when required by a HART digital communicator and avoids the inconvenience of carrying a 250 O resistor
|
IDOS™ Flexibility
Intelligent Digital Output Sensor (IDOS)
|
Universal Pressure Modules are available from 10 in H2O to 10,000 psi (25 mbar to 700 bar).
|
Total Flexibility
|
IDOS modules can be used with any compatible instrument; for example, a DPI 812 RTD loop calibrator can become a fully featured pressure calibrator.
|
Plug and Play
|
Modules are interchangeable between instruments, requiring no set-up or instrument calibration.
|
Range Expansion
|
Achieved by adding modules (please refer to the IDOS Universal Pressure Module datasheet).
|
*S_Standard Accuracy
|
Total accuracy over 32°F to 122°F (0°C to 50°C), including one year stability and calibration uncertainty
|
*P_Premier Accuracy
|
Precision over 65°F to 82°F (18°C to 28°C)
For operation over 41°F to 113°F (5°C to 45°C): 0.014% FS for ranges above 10 psi (700 mbar) 0.075% FS for ranges below 15 psi (1 bar)
Stability over a year: 0.01% reading ranges above 5 psi (350 mbar) 0.03% reading ranges below 10 psi (700 mbar)
Calibration uncertainty: 50 ppm of reading
|
Single or DualRange
|
One or two internal sensors can be selected. For dual range instruments, G/D ranges will be configured as G (gauge).
Only one of the two sensors can be 1500 psi (100 bar) or above.
|
Overpressure (maximum transient/intermittent pressure)
|
5 psi (350 mbar) and below 4 x FS
10 to 10,000 psi (700 mbar to 700 bar) 2 x FS
Maximum working pressure: 1.1 x FS
|
Pressure Connections
|
1/8 NPT female or G 1/8 female
|
DPI 800 Series Common Specification
Operating Temperature
|
14°F to 122°F (-10°C to 50°C)
|
Storage Temperature
|
-4°F to 158°F (-20°C to 70°C)
|
Humidity
|
0% to 90% non-condensing, Def Stan 66-31, 8.6 Cat III
|
Shock and Vibration
|
BS EN61010:2001, Def Stan 66-31, 8.4 Cat III
|
EMC
|
BS EN61326-1:1998 + A2:2001
|
Safety
|
Electrical BS EN61010:2001. Pressure Equipment Directive (PED), Class SEP. CE marked
|
Display
|
Graphic LCD with backlight. Resolution 99999
|
Size (l x w x h) and Weight
|
7.1 in x 3.3 in x 2 in (180 mm x 85 mm x 50 mm), 18 oz (500 g)
|
Batteries
|
3 AA alkaline, >50 hours measure, >10 hours 24V source
|
|