- Ranges:
+ Water 20 °C, min: 0,25-2,5 l/h, max:16-160 l/h
+ Air 20 °C, 1,2 bar abs., min: 0,5-5 Nl/h, max:500-5.000 Nl/h
- Electronic output (Option): inductive limit switches
- Accuracy:
+ Liquids / gases: ± 2,5% q G 50 acc. VDE/VDI 3513 page 2
+ K12-V / N (differential pressure regulator) ± 3 % / ± 5 % FS within 10-100% of the range
- Repeatability Liquids / gases ± 1,0 % FS
+ K12-V / N (differential pressure regulator) ± 1,5 % / 2,5 % FS
- Applications:
+ Metering of liquid products with suitable viscosity, free of solids, not clog and do not tend to form deposits.
+ Gases with linear flow characteristics and adequate inlet pressure.
- Ambient conditions:
+ Ambient temperature -20°C up to +100°C (risk of frost fracture!)(-20°C up to +70°C)
- Product conditions :
+ without switch: -20°C up to +100°C (with switch: -20°C up to +70°C)
- Ingress protection: IP 65 (EN60529)