TECHNICAL SPECIFICATIONS HYPERLAB PLUS and BASIC
VOLUMES (Hyperlab Plus and Basic)
-Sample: 2-300µl, precision 1% at 10µl
-Reagent: 5-350 µl, precision < 1% at 250µl
-Reaction (sample + reagent): 220-350µl
-Mixing in cuvette after dispensation
DILUTION (Hyperlab Plus and Basic)
-Automatic programmable dilution for each sample
-Factors: 1:1 1:4 1:10 1:40 1:100
ANALYTICAL SPEED
-125...200 analyses/hour - Hyperlab Plus
-100...180 analyses/hour - Hyperlab Basic
TEMPERATURE CONTROL (Hyperlab Plus and Basic)
- Refrigerated reagent at 15°C below ambient temperature
- Reading cuvette with programmable heating
- Reagent preheating in sucking needle
TYPES OF TESTS (Hyperlab Plus and Basic)
Endpoint, Bichromatic endpoint, Differential endpoint, Differential endpoint sample blank, Derivatives, Fixed time, Kinetic.
SAMPLES TRAY
-60 samples positions - Hyperlab Plus
-30 vials and 20 cups - Hyperlab Basic
REAGENTS TRAY
-30 positions refrigerated tray - Hyperlab Plus
-26 positions refrigerated tray - Hyperlab Basic
SAMPLING NEEDLE
- Glass ceramics stroke needle, capacitance sensor, reagent pre-heating - Hyperlab Plus
- Stainless steel needle, capacitance sensor, optional reagent pre-heating - Hyperlab Basic
REACTION CUVETTE ROTOR (Hyperlab Plus and Basic)
80 BIONEX® cuvettes replaceable singularly, over 20.000 tests
WASHING STATION
-6 dispensing needles for washing solution, 6 sucking needles, 1 drying needle, 8 washing cycles for each cuvette - Hyperlab Plus
-4 dispensing needles for washing solution, 5 sucking needles, 1 drying needle, 6 washing cycles for each cuvette - Hyperlab Basic
|