Size and Weight: |
85mm x 255mm x 46mm – 600 grams |
Measuring Principle: |
Infrared spectrometry ethanol molecule-selective 9.46 µm-wavelength |
Measuring Cycle: |
Automatic with monitoring and diagnosis of good operation |
Measurement Unit: |
mg/L of alcohol in exhaled air, µg/L air, g/L blood BAC, %BAC |
Measuring Range: |
0 to 2,00 mg/L (0 to 450 BAC) |
Measurement Display: |
Function of unit chosen, accuracy maximum 1 µg/L |
Measuring Frequency: |
Ready to blow in less than 30 seconds |
Calibration: |
Dry gas / wet gas |
Operating Temperature: |
-5°C to +40°C |
Data Transfer: |
PC via USB port and ACS software |
Memory: |
Over 3,200 time-stamped events with Bluetooth printing |
Display: |
Graphic LCD |
Power Supply: |
Built-in rechargeable NiMH battery |
Features: |
Bluetooth printing and GPS applications |
Battery Life: |
Over 20 hours |
Certification: |
OIML R 126 1998 – edition international recommendation |