Clear-view cooling thermostat with insulating anti-fog double-pane glass window ideally suited for standard viscosity test of liquids by means of glass capillary viscometers according to GOST 33, GOST R 53708, ASTM D445, IP 71, ISO 3104 and DIN 51366.
Features & Benefits
Thanks to the built-in refrigeration machine, CRYO‑VIS‑T‑03 lets you proceed viscosity measure at temperatures below ambient without using an external chiller.
The innovative design of the temperature control unit, which unites a sensor, a heater and a pump in one chamber, achieves exceptional stability and uniformity of the temperature field in the working volume.
Optimal control of the refrigeration machine ensures rapid heating and cooling of the thermostat when changing the set point.
Floating guide with lock for vertical installation of viscometer.
Holders with spring clamp provide fast and reliable attachment of viscometer with tube of any diameter.
Exceptional visibility for precise measurement due to backlight lamp.
Rotating feet to install the bath on a horizontal surface.
Intuitive operation and bright indicators "Stable state" and "Alarm".
Three programmable set point temperatures.
Pump made of 304 stainless steel ensures a long lifetime with any bath fluid in a wide temperature range.
Built-in clock and stopwatch for timing during measurement.
Switching on and off at a given time due to the built-in clock.
Digitally adjustable high temperature cut-off, heater failure and dry-running protection guarantee absolute safety of operation even without supervision.
Built-in USB and optional RS232/RS485 interfaces to control the device in open protocol.
Thông số kỹ thuật
Working temperature range |
|
|
|
|
|
Recommended thermal fluid |
|
- for temperature range +5...+50 °С
|
|
- for temperature range -30...+50 °С
|
|
|
|
Cooling capacity at -30 °С |
|
|
|
|
|
|
|
|