Model |
VM-3004 SI |
Frequency Range |
10Hz~1000Hz |
Full scale |
Acceleration (ACC): 1, 10, 100 m/s2 3-ranges,Min 1 digit : 0.01 m/s2
Velocity (VEL): 1, 10, 100 mm/s 3-ranges,Min 1 digit : 0.001 mm/s
Displacement (DISP): 10, 100, 1000 μm p-p 3-ranges,Min 1 digit : 0.01μm p-p |
Indication |
Equivalent-peak, rms for Acceleration and Velocity,
Equivalent peak for Displacement |
Accuracy |
Frequency Response
|
±5%(20~500Hz)+5% -15%(10Hz~1000Hz) |
Sensitivity Error |
±5% (at 80Hz) |
Range Changeover Error |
±2% (at 80Hz) |
Linearity |
±0.5% of F.S.(at 80Hz) |
Level difference Input- Output of Frequency Analysis (for Full Scale) |
±5% of F.S. |
Frequency Analysis Selectivity |
Q≒10 |
Frequency Analysis Indication Error |
±5% |
Output |
- Waveform output
- Level output
- TUNE OUT
|
AC OUT 0 ~ ±1V (with a load≧10kΩ)
DC OUT 0 ~ +1V (with a load≧10kΩ)
0.1~1V/10~100Hz or 100~1000Hz |
Power Supply |
Battery 6F22:1pce, Life time:16hrs approx.at continuous use
External battery AC Power supply 9 ~ 18 V, 20mA max. |
Calibration Output |
80Hz sinusoidal wave output
Level is adjustable using variable resistor in battery box |
Dimensions & Weight |
Main unit:80Wx160Hx36Dmm, 380g |
Pickup Dimensions & Weight |
Pickup: 28Wx70Hx38Dmm, 165g
Probe: φ8x62 mm, 33g |
|