Gas Measured: |
Carbon Dioxide (CO2), and Oxygen (O2) |
Measurement Methods: |
CO2: NDIR (Non-Dispersive Infra-Red.)
O2: Electro Chemical Sensor |
Measuring Ranges: |
CO2: 00.00-100.0%
O2: 00.000-100.0% |
Resolution: |
CO2: 0.01% Below 10.0%, 0.1% above 10.0%,
O2: 0.001% Below 2.00%, 0.01% above 2.00% |
Accuracy |
4% relative to gas reading |
Repeatability |
1% relative to gas reading |
Response time: |
8 seconds to 95% Final Reading,
15 Seconds to 99% Final Reading. |
Sampling Rate |
4 ml/sec |
Sample Time |
Factory default 15 Seconds. Adjustable in 1 second increments. |
Gas-borne Contaminate Protection Built-In |
External and internal water blocking and particle filters protect analyzer from test gas-borne contamination. |
Warm-up: |
Full accuracy in 5 minutes. |