Radiation detected |
Gamma & X-ray |
Energy threshold |
40 keV |
Detector |
Silicon semiconductor |
Measurement range |
Dose: 1 µSv to 10 SvDose rate: 1 µSv to 15 Sv |
Accuracy |
Within ±10% from 10 µSv/h to 10 Sv/hCalibrated by 137Cs using a slab phantom |
Linearity |
Within ±20% from 10µSv/h to 1Sv/h |
Energy dependence |
Within ±30% from 50 keV to 1.5 MeV |
Display |
4-dgit LCD with unit and battery indications |
Switch |
Power on/off (including reset operation) and dose rate display |
Immunity |
150 V/m : 0.15 MHz to 200 MHz100 V/m : 200 MHz |
Power |
One lithium battery CR2450 (conforming to IEC 60086-2) |
Environment (operational) |
Temp -10°C to +50°CRelative humidity up to 90% (non-condensing) |
Dimensions |
Approx. 30 x 11 x 108mm (excluding the clip) |
Weight |
Approx 40g |