For laboratory use
For low-pressure measurement up to 2 kPa
Analogue output, Mini2Logger and RS232 connections
Integrated barometer module<
Duct cross-section input facility per keystroke (for l/min or m3/h)
Flow calculation based on standard or variable density
Instrument specifications:
2 Models:
measuring range ±200 Pa
measuring range ±2 kPa
Measurement units with changeover facility:
Pressure: Pa, mbar, mmH2O
Flow: m/s, l/min, m3/h
Barometer: kPa, bar
with humidity probe: % rF
with themp. probe: ºC
Minimum, mean and maximum values per keystroke
Automatic temperature compensation of pressure sensor
Connection facility for external humidity and temperature probes
Thermocouples type E, J, K and T can be used in addition to NTC- and PT1000 temperature probes
Temperature, humidity and barometer readings available per keystroke
Thông số kỹ thuật
|
Measuring range |
Resolution |
Accuracy |
ManoAir500 Mod. 1
100 % range
10 % range |
±2000 Pa
±20 Pa |
0,1 Pa
0,01 Pa |
±0,2 Pa
±0,05 Pa |
ManoAir500 Mod. 2
100 % range
10 % range |
±2000 Pa
±20 Pa |
1 Pa
0,1 Pa |
±2 Pa
±0,5 Pa |
Barometer (static pressure) |
0 … 400 kPa (4 bar) |
1 kPa |
±1 kPa |
Humidity probe
Humidity::
Temperature: |
0 … 99 % rH
−20 … 60 °C |
0,1 % rH
0,1 °C |
±3 % rH
±0,5 °C |
Universal-temperature |
−20 … 140 °C |
0,1 °C |
≥0≤70 °C: 0,2 °C
>70 °C: 0,5 °C |
Air-temperature probe |
−20 … 140 °C |
0,1 °C |
≥0≤70 °C: 0,2 °C
>70 °C: 0,5 °C |
Surface-temperature probe |
−20 … 140 °C |
±0,1 °C |
≥0≤70 °C: 0,2 °C
>70 °C: 0,5 °C |
Universal high temperature probe |
−20 … 600 °C |
0,1 °C |
≥0≤70 °C: 0,5 °C
>70 °C: 1 °C |
Thermocouple with adapter |
depending on applied element |
1 °C |
depending on applied element |